10
Setup screens
Stability criteria
Options:
Accurate
,
Medium
,
Fast
.
Press the corresponding functional key to select the desired option.
Select
Accurate
for accurate measurements; long duration.
Select
Medium
for normal measurements.
Select
Fast
only for quick estimations; low accuracy in short time.
While in the
Setup
menu it is possible to modify the instrument’s parameters. By pressing the
/ keys it is possible to move between the
Setup
options. Each menu item displays the
current option. Press
HELP
to view the contextual help.
Temperature compensation
Options:
Enable, Disable
Enable this feature only if
HI 7662-T
temperature probe is
connected and electrode isopotential point is known.
If enabled, the meter will take into account the temperature of
the readings when evaluating the final result.
If disabled, the computed concentration will not be temperature
compensated.
Press the functional key to select the desired option.
Isopotential point
Options:
1.0 to 50.0 mg/L (ppm).
Press
Modify
to access the isopotential point concentration
modify screen.
Press / to increase/decrease the isopotential point concentration.
Press
Accept
to confirm the value or
ESC
to return to the
Setup
menu without saving.
If the temperature compensation feature is disabled the
meter will display
“None”
for the isopotential point, and the
isopotential point modify screen cannot be accessed.
Concentration unit
Options:
g/L, ppt
Press the corresponding functional key to select the desired
concentration unit to be used in displaying measurements results.