37

Dissolved Oxygen

• Let the sample stand and the flocculent agent will start to settle.

• After approximately 2 minutes, when the upper half of the bottle becomes

limpid, add 10 drops of HI 93732C-0.

• Replace the cap and invert the bottle until the settled flocculent dissolves completely. The sample is

ready for measurement when it is yellow and completely limpid.

• Fill the cuvette up to the mark with 10 mL of the unreacted (original)

sample, and replace the cap. This is the blank.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when

the meter is zeroed and ready for measurement.

• Remove the cuvette.

• Fill another cuvette up to the mark with 10 mL of the reacted sample and

replace the cap.

• Reinsert the cuvette into the instrument.

• Press READ to start the reading. The instrument will display the results in mg/L of dissolved oxygen.

10 mL

x 10

INTERFERENCES

Interferences may be caused by reducing and oxidizing materials.

Dissolved Oxygen

SPECIFICATIONS

Range

0.0 to 10.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.4 mg/L ±3% of reading

Typical EMC

± 0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the 

Standard Methods for the Examination of Water and Wastewater,

18

th

 edition

,

 

Azide modified Winkler method. The reaction between dissolved oxygen

and the reagents causes a yellow tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI  93732

A

-0

Reagent A

5 drops

HI  93732

B

-0

Reagent B

5 drops

HI  93732

C

-0

Reagent C

10 drops

REAGENT SET

HI 93732-01

Reagents for 100 tests

HI 93732-03

Reagents for 300 tests

For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the 

Dissolved Oxygen

 method using the procedure de-

scribed in the 

Method Selection

 section (see page 11).

• Fill one 60 mL glass bottle completely with the unreacted sample.

• Replace the cap and ensure that a small part of the sample spills

over.

• Remove the cap and add 5 drops of HI 93732A-0 and 5 drops of

HI 93732B-0.

• Add more sample, to fill the bottle completely. Replace the cap

again and ensure that a part of the sample spills over. This is to
make sure that no air bubbles have been trapped inside, which
could alter the reading.

• Invert several times the bottle. The sample becomes orange-yellow

and a flocculent agent will appear.

x 5

x 5

DISSOLVED OXYGEN

Summary of Contents for HI 83203

Page 1: ...1 Instruction Manual HI 83203 Multiparameter Bench Photometer for Aquaculture w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Page 2: ...tometer can be connected to a PC via an USB cable The optional HI 92000 Windows Compatible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please rea...

Page 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Page 4: ...creen NEED TO KNOW HI 83203 has a powerful interactive user support that assists the user during the analysis process Each step in the measurement process is help supported A tutorial mode is availabl...

Page 5: ...a b a c a b shake gently shakevigorously b c USING CUVETTES Proper mixing of the cuvette is done by shaking the cuvette moving the cuvette up and down The movement may be gentle or vigorous This mixi...

Page 6: ...before performing tests Safety equipment Wear suitable eye protection and clothing when required and follow instructions carefully Reagent spills If a reagent spill occurs wipe up immediately and rins...

Page 7: ...tothe setup menu without saving the new value SETUP SETUP SETUP SETUP SETUP DATA MANAGEMENT The instrument features a data log function to help you keep track of all your analysis The data log can hol...

Page 8: ...ument serial number Press ESC to return to the Setup mode HELP MODE HELP MODE HELP MODE HELP MODE HELP MODE HELP MODE Date Time This option is used to set the instrument s date and time Press Modify f...

Page 9: ...arious aliphatic and aromatic amines Ammonia MR x 4 x 4 Ammonia MR SPECIFICATIONS Range 0 00 to 10 00 mg L Resolution 0 01 mg L Accuracy 0 05 mg L 5 of reading Typical EMC 0 01 mg L Deviation Light So...

Page 10: ...amines sulfide various aliphatic and aromatic amines x 4 x 4 Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Ligh...

Page 11: ...ke gently again x 3 x 3 FREE CHLORINE FREE CHLORINE FREE CHLORINE FREE CHLORINE FREE CHLORINE SPECIFICATIONS Range 0 00 to 2 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0...

Page 12: ...ests powder For other accessories see page 46 MEASUREMENT PROCEDURE Select the Total Chlorine method using the proce dure described in the Method Selection section see page 11 Fill the cuvette with 10...

Page 13: ...partially or may rapidly fade To resolve this neutralize the sample with diluted HCl or NaOH Total Chlorine Total Chlorine Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for...

Page 14: ...5 00 mg L Resolution 0 01 mg L Accuracy 0 02 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the EPA meth...

Page 15: ...0 to 1000 g L Resolution 1 g L Accuracy 10 g L 5 of reading Typical EMC 1 g L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the EPA method The r...

Page 16: ...amines Chloride above 100 ppm Chlorine above 2 ppm Copper Iron III Strong oxidizing and reducing substances Sulfide must be absent Nitrate SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 1 mg L Ac...

Page 17: ...sten lamp with narrow band interference filter 575 nm Method Adaptation of the Ferrous Sulfate method The reaction between nitrite and the reagent causes a greenish brown tint in the sample REQUIRED R...

Page 18: ...Nitrite LR SPECIFICATIONS Range 0 00 to 0 35 mg L Resolution 0 01 mg L Accuracy 0 02 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter...

Page 19: ...ading Typical EMC 0 1 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 420 nm Method Adaptation of the Standard Methods for the Examination of Water and Wastewater 18th e...

Page 20: ...scription Quantity HI 93710 0 Phenol Red Indicator 5 drops REAGENT SETS HI 93710 01 Reagents for 100 pH tests HI 93710 03 Reagents for 300 pH tests For other accessories see page 46 MEASUREMENT PROCED...

Page 21: ...RANGE SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 1 mg L Accuracy 1 mg L 4 of reading Typical EMC Dev 0 1 mg L Light Source Tungsten lamp with narrow band interference filter 525 nm Method Ada...

Page 22: ...cy 0 04 mg L 4 of reading Typical EMC Dev 0 01 mg L Light Source Tungsten lamp with narrow band interference filter 610 nm Method Adaptation of the Ascorbic Acid method The reaction between phosphate...

Page 23: ...ETHODS The analyzed data can be managed using Hanna s product HI 92000 Windows Compatible Software DATA MANAGEMENT Theinstrumentshowsclearwarningmessageswhenerroneousconditionsappearandwhenmeasuredval...

Page 24: ...instruments may cause unacceptable interferences to other electronic equipments this requiring the operator to take all necessary steps to correct interferences Any variation introduced by the user to...

Page 25: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Reviews: