23
Calibration
Using DO standards
The meter can be calibrated in a single or multi-points (up to two points), using 2 DO standards (0.0%, 100%)
or using the custom standards.
The following options are available for calibration:
Standard Recognition
The user can choose between
Automatic recognition (from 2 Hanna standards available) or User Standard.
• Use
or
to highlight the
Standard Recognition
option.
• Use
or
to choose from
Automatic or User
Standard option.
• Press
to confirm your selection or press
to
cancel operation.
Calibration Reminder
This option allows the user to set the calibration reminder as
Daily, Periodic or Disabled.
To set the calibration reminder:
• Use
or
to highlight the
Calibration re-
minder option.
• Press
to confirm your selection and then use
or
to choose the desired option
.
• Press
to confirm your selection or press
to
cancel operation.