2
HI 4007 Chloride Half-cell
HI 4007 Chloride Half-cell
HI 4007 Chloride Half-cell
HI 4007 Chloride Half-cell
HI 4007 Chloride Half-cell
HI 4107 Chloride Combination Electrode
HI 4107 Chloride Combination Electrode
HI 4107 Chloride Combination Electrode
HI 4107 Chloride Combination Electrode
HI 4107 Chloride Combination Electrode
I.
I.
I.
I.
I. Introduction:
Introduction:
Introduction:
Introduction:
Introduction:
The Hanna HI 4007and HI 4107 are ion selective elec-
trodes designed for the measurement of chloride ions in
aqueous solutions. The HI 4007 is a solid state half-cell
sensor that requires a separate reference. The HI 4107 is a
combination ion selective electrode.
II
IIII
IIII.
Specifications
Specifications
Specifications
Specifications
Specifications
Type:
Solid State electrode with
a Silver Chloride pellet.
Ion(s) measured:
Chloride (Cl
-
)
Measurement range: 1.0 M to 5 X 10
-5
M
35,500 to 1.8 ppm
Interfering ions:
Cyanide, and Mercury ions
must be absent. Ratio of interfering ion to Cl
-
must be less
than the ratio indicated below:
1.0
for l
-
iodide
3.5
for Br
-
bromide
for CO
3
2-
carbonate
1.0
for OH
-
hydroxide
0.01
for S
2
O
3
2-
thiosulfate
Operating Temperature:
0-80
°
C
Operating pH:
2-11 pH
Dimensions:
12 mm (OD) X 120 mm
nominal insertion
(0.47
”
X 4.72
”
)
Connection:
BNC