7
HI 4007
•
Remove protective cover from sensor tip.
•
Prepare HI 5315 reference electrode by filling electro-
lyte reservoir with HI 7072 fill solution.
•
Place sensor and reference electrodes into electrode
holder and connect cable connectors to meter.
HI 4107
•
Remove the protective plastic wrap that covers the
ceramic junction before assembling sensor for the first
time.
•
HI 7072 reference fill solution should be added daily
to electrolyte reservior before electrode use.
•
During measurement always operate electrode with
the fill hole open.
•
During normal use, fill solution will slowly drain out
of the tapered cone junction at the lower portion of the
electrode. Excessive loss (>4 cm drop within 24 hours)
is not normal. If this occurs verify cap is tightened and
the interface between the internal cone and outer
body is free of debris.
•
Add fill solution daily to maintain a good head pres-
sure. For optimum response, this level should be main-
tained and not be allowed to drop more than 2-3 cm
(1-inch) below fill hole. It must cover the ceramic
found on the inner stem.
•
If an erratic measurement occurs, check to see if for-
eign matter is seen trapped near the internal cone.
Drain and refill with fresh fill solution.
•
Avoid large changes in temperature (thermal shock)
as it may damage the sensor.