23
22
“3” for 1 minute
“4” for 5 minutes
“5” for 30 minutes
“6” for 60 minutes
“7” for 120 minutes
“8” for 180 minutes
14) last sample time, min. E.g. “59” for 59 minute
15) last sample time, hour. E.g. “12” for 12 hour
16) last sample time, day. E.g. “09” for 9th day
17) last sample time, month. E.g. “09” for September
18) last sample time, year. E.g. “96” for year 1996
19) etx end
?DM
Requests the meter to send the selected lot data memory. The
data is sent in the following order:
1) stx
2) Lot number E.g. “01” for lot No. 1
Note:
From 3) to 12) steps see the 4) to 13) steps of “
?VM
” command.
13) Total number of samples per lot
E.g. “1234” for total no. of samples: 1234.
14) Logged data in signed integer, repeat sending in logged
channel sequence
sample no. 1
—— send ppm data if ppm is log selected
—— send % data if % is log selected
—— send temperature data if temperature is log selected
sample no. 2 ...
until the last sample
sample “XXX” is signed hex format.
15) last sample time, min. E.g. “59” for 59 minute
16) last sample time, hour. E.g. “12” for 12 hour
17) last sample time, day. E.g. “09” for 9th day
18) last sample time, month. E.g. “09” for September
19) last sample time, year. E.g. “96” for year 1996
20) etx end
Commands setting parameters:
/ML
To select the data lot for data transfer.