13
12
S E T U P
Setup mode allows viewing and modifying the following parameters:
• Salinity Compensation
• Altitude Compensation
• Log Interval
• Time
• Date
• Baud Rate
• Command Prefix
• Beep Status
To enter the Setup mode press
SETUP
while the instrument is in measurement
mode. Press
SETUP
again to exit SETUP mode.
Select a parameter with the
ARROW
keys.
Press
CAL
if you want to change a parameter value. The selected parameter
will blink.
Press the
ARROW
keys to increase or decrease the displayed value.
Press
CFM
to save the modified value or
CAL
to escape.
Press the
ARROW
keys to select the next/previous parameter.
S
ALINITY COMPENSATION SET
Press
CAL
when the salinity factor is displayed. The salinity factor (0 to 40
g/l) will start blinking.
Press
ARROW
keys to change the salinity factor value.
Press
CFM
to save the modified value or press
CAL
to escape without saving.
CFM
• Immerse the D.O. probe in the second vessel as
near as possible to the
ChecktempC
. Allow a
few seconds for the probe to stabilize.
• Use the " " or " " key to set the reading on
the secondary LCD to that of the hot water.
• When the reading is stable, "CFM" tag starts
blinking.
• Press
CFM
to confirm. The instrument returns to
measurement mode.
Note:
If the reading is not close to the selected calibration point, "WRONG"
tag will blink. In this case, check if the value set on the secondary LCD
and the temperature measured with
ChecktempC
are close. Change
the D.O. probe or restart calibration if necessary.
NaCl
g / l