49
KDS
null
Same as LCD key
KDW
null
Same as key
KRG
null
Same as key
KST
null
Same as SETUP key
KUP
null
Same as key
MDR
null
Request firmware code
ECR
null
Request EC reading (available in
control or
idle mode only)
TDR
null
Request TDS reading (available
in control
or idle mode only, for
HI 24
series only)
RNG
null
Request measure Range
(available in
control
or idle mode only)
TMR
null
Request temperature reading
PWD
NNNN
Send the 4-digit password
SET
NNPC
1
C
2
C
3
C
4
C
5
Set setup item NN to the PC
1
C
2
C
3
C
4
C
5
value; P=+ if the value is greater than 0, P=- if the value
is less than 0. C
1
can be 0 or 1; only C
2
C
3
C
4
C
5
can be 0 to 9
or blank (the command is not available if the controller is in
setup mode)
Note
If the controller is not in control or idle mode and the tem-
perature reading is requested through the TMR command,
the controller answers with the last acquired reading when it
was in control or idle mode.
Following are examples of commands for setup items:
“01200<CR>”
“01200<CR>”
“01200<CR>”
“01200<CR>”
“01200<CR>”
This command sets the setup item 22 (relay 2 set point) of a
EC controller, identified by the process ID number 03, to the
+12.00 mS value.