36
CLEANING PROCEDURE
• General
Soak in Hanna Instruments
HI7061
or
HI8061
General
Cleaning Solution for approximately ½ hour.
• Protein
Soak in Hanna Instruments
HI7073
or
HI8073
Protein
Cleaning Solution for 15 minutes.
• Inorganic
Soak in Hanna Instruments
HI7074
Inorganic Cleaning
Solution for 15 minutes.
• Oil/grease
Rinse with Hanna Instruments
HI7077
or
HI8077
Oil
and Fat Cleaning Solution.
IMPORTANT:
After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel‑filled electrodes) and soak
the electrode in
HI70300
or
HI80300
Storage Solution for at least
1 hour before taking measurements.