pH BUFFER TEMPERATURE
DEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions are
affected by temperature changes to a lesser degree than normal solutions.
During calibration the instrument will automatically calibrate to the pH
value corresponding to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25 ºC.
TEMP
pH BUFFERS
°C °F 1.68 4.01 6.86 7.01 9.18 10.01 12.45
0
32
1.67 4.01 6.98 7.13 9.46 10.32 13.38
5
41
1.67 4.00 6.95 7.10 9.39 10.24 13.18
10 50
1.67 4.00 6.92 7.07 9.33 10.18 12.99
15 59
1.67 4.00 6.90 7.05 9.27 10.12 12.80
20 68
1.68 4.00 6.88 7.03 9.22 10.06 12.62
25 77
1.68 4.01 6.86 7.01 9.18 10.01 12.45
30 86
1.68 4.02 6.85 7.00 9.14 9.96 12.29
35 95
1.69 4.03 6.84 6.99 9.11 9.92 12.13
40 104
1.69 4.04 6.84 6.98 9.07 9.88 11.98
45 113
1.70 4.05 6.83 6.98 9.04 9.85 11.83
50 122
1.71 4.06 6.83 6.98 9.01 9.82 11.70
55 131
1.72 4.08 6.84 6.98 8.99 9.79 11.57
60 140
1.72 4.09 6.84 6.98 8.97 9.77 11.44
65 149
1.73 4.11 6.84 6.99 8.95 9.76 11.32
70 158
1.74 4.12 6.85 6.99 8.93 9.75 11.21
75 167
1.76 4.14 6.86 7.00 8.91 9.74 11.10
80 176
1.77 4.16 6.87 7.01 8.89 9.74 11.00
85 185
1.78 4.17 6.87 7.02 8.87 9.74 10.91
90 194
1.79 4.19 6.88 7.03 8.85 9.75 10.82
95 203
1.81 4.20 6.89 7.04 8.83 9.76 10.73