• When the reading is stable and close to the selected calibration
point, “
READY
” tag will appear and “
CFM
” tag will blink.
• Press
CFM
to confirm. The instrument returns to measurement mode.
Note: If the reading is not close to the selected calibration point,
“WRONG” tag will blink. Change the temperature probe and restart
calibration.