background image

11

10

CONDUCTIVITY PROBE  MAINTENANCE

Rinse the probe with tap water after
every series of measurements.
If a more thorough cleaning is re-
quired, remove the sleeve and clean
the probe with a cloth or a non-
abrasive detergent.
After cleaning, re-calibrate the in-
strument.
The 4-platinum-ring probe body is
made with glass. For this reason,
great care must be taken while
handling the probe.

• Immerse the probe into a sample of the solution and adjust  the

conductivity TC knob to 0% (i.e. no compensation).

• Condition sample and probe at 25°C, and note the conductivity

reading C

25

.

• Condition sample and probe to a temperature T, approximately

5°C to 10°C different from 25°C, and note the conductivity
reading C

T

.

• The temperature coefficient 

β

 of the solution is calculated with the

following formula: 

β

= 100 x (C

T

-C

25

) / [(T - 25) x C

25

]

The above procedure is suitable for determining the temperature
coefficient in the laboratory where the temperature of the solution can
be determined and controlled. If this is not possible, e.g. during field
measurements, the following procedure should be used:
• Immerse the probe into the test solution and turn the conductiv-

ity TC knob to 0% (no compensation).

• Allow the reading to stabilize (the reading should not change by

more than ±0.2 mS within 1 minute) and record the value, C.

• Repeat the procedure with the temperature of the solution

changed by more than 10°C. Wait for the conductivity reading to
stabilize.

• Adjust the conductivity TC knob until the display shows the

previously recorded C value.

• The value indicated by the knob is the temperature coefficient of

the solution.

DETERMINATION OF THE TEMPERATURE

COEFFICIENT  OF  A  SOLUTION

Summary of Contents for HI 216

Page 1: ...HI 216 EC Resistivity Meter Instruction Manual w w w h a n n a i n s t c o m...

Page 2: ...the rear panel and the meter automatically recognizes which probe is connected GENERAL DESCRIPTION Dear Customer Thank you for choosing a Hanna Instruments Product Please read this instruction manual...

Page 3: ...bes included HI 76303 for conductivity measurements HI 3316D for resistivity measurements Power Supply 12 Vdc power adapter included Environment 0 to 50 C 32 to 122 F RH max 95 Dimensions 240 x 182 x...

Page 4: ...ity TC knob to the tem perature coefficient value of the solution see Determination of the temperature coefficient of a solution section for details Select the appropriate conductivity range by pressi...

Page 5: ...be referenced to 25 C 77 F If you need to reference the measurements to 20 C 68 F adjust the calibration knob to read the calibration solution value 20 C See conductivity vs temperature chart Calibra...

Page 6: ...ient of the solution is calculated with the following formula 100 x CT C25 T 25 x C25 The above procedure is suitable for determining the temperature coefficient in the laboratory where the temperatur...

Page 7: ...HI 7031M 1413 S cm 230 mL bottle HI 7031L 1413 S cm 500 mL bottle HI 8031L 1413 S cm 500 mL FDA bottle HI 7033M 84 S cm 230 mL bottle HI 7033L 84 S cm 500 mL bottle HI 8033L 84 S cm 500 mL FDA bottle...

Page 8: ...are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions The probes are guaranteed for a period of six mo...

Page 9: ...Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011...

Reviews: