PRELIMINARY  EXAMINATION

HI 216 

is a combination bench meter that can read conductivity in 4

different ranges, and resistivity.
For conductivity measurements, the calibration is a simple 1 point
procedure through the easy-to-operate front panel knob, and the
supplied probe does not require re-calibration when switching from
one range to another.
The 4-platinum-ring probe has a built-in temperature sensor that
automatically compensates for temperature changes, and the tem-
perature coefficient can be adjusted from 0 to 2.5% through a knob
on the front panel.
For resistivity measurements, the meter is factory calibrated and, if
necessary, calibration can be adjusted by acting on the trimmer
located on the rear panel.
The 

HI 3316D

 resistivity probe is easy to clean and requires little

maintenance. It also features a built-in temperature sensor to
automatically compensate for temperature variations, with an user-
selectable (from 2 to more than 7%) temperature coefficient.
Both probes use the same DIN plug on the rear panel, and the meter
automatically recognizes which probe is connected.

GENERAL  DESCRIPTION

Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instru-
ment. This manual will provide you with all the necessary information
for correct operation.
If you need additional technical information, do not hesitate to e-mail
us at 

[email protected]

.

This instrument is in compliance with the 

 directives.

PRELIMINARY EXAMINATION ................................................................... 3

GENERAL DESCRIPTION ......................................................................... 3

FUNCTIONAL DESCRIPTION ..................................................................... 4

SPECIFICATIONS .................................................................................... 5

OPERATIONAL GUIDE ............................................................................. 6

CALIBRATION ....................................................................................... 8

DETERMINATION OF THE TEMPERATURE COEFFICIENT ............................ 10

CONDUCTIVITY PROBE MAINTENANCE ................................................... 11

ACCESSORIES ..................................................................................... 12

WARRANTY ........................................................................................ 14

Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your dealer.
Each meter is supplied complete with:

• HI 76303 

conductivity probe

HI 3316D

 resistivity probe

12 Vdc adapter

Instruction manual

Note:

Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the
original packing with the supplied accessories.

TABLE OF CONTENTS

Summary of Contents for HI 216

Page 1: ...HI 216 EC Resistivity Meter Instruction Manual w w w h a n n a i n s t c o m...

Page 2: ...the rear panel and the meter automatically recognizes which probe is connected GENERAL DESCRIPTION Dear Customer Thank you for choosing a Hanna Instruments Product Please read this instruction manual...

Page 3: ...bes included HI 76303 for conductivity measurements HI 3316D for resistivity measurements Power Supply 12 Vdc power adapter included Environment 0 to 50 C 32 to 122 F RH max 95 Dimensions 240 x 182 x...

Page 4: ...ity TC knob to the tem perature coefficient value of the solution see Determination of the temperature coefficient of a solution section for details Select the appropriate conductivity range by pressi...

Page 5: ...be referenced to 25 C 77 F If you need to reference the measurements to 20 C 68 F adjust the calibration knob to read the calibration solution value 20 C See conductivity vs temperature chart Calibra...

Page 6: ...ient of the solution is calculated with the following formula 100 x CT C25 T 25 x C25 The above procedure is suitable for determining the temperature coefficient in the laboratory where the temperatur...

Page 7: ...HI 7031M 1413 S cm 230 mL bottle HI 7031L 1413 S cm 500 mL bottle HI 8031L 1413 S cm 500 mL FDA bottle HI 7033M 84 S cm 230 mL bottle HI 7033L 84 S cm 500 mL bottle HI 8033L 84 S cm 500 mL FDA bottle...

Page 8: ...are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions The probes are guaranteed for a period of six mo...

Page 9: ...Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011...

Reviews: