5
4
SPECIFICATIONS
Range
199.9 μS/cm / 0 to 1999 μS/cm
19.99 mS/cm / 199.9 mS/cm
0 to 19.90 M
Ω
•
cm
Resolution
0.1 μS/cm / 1 μS/cm
0.01 mS/cm / 0.1 mS/cm
0.10 M
Ω
•
cm
Accuracy (@20°C/68°F)
±1% FS for EC
±2% FS for resistivity
Calibration
Manual, 1 point, for both EC and resistivity
Temperature Compensation
Automatic from 0 to 50°C
with ß selectable from 0 to 2.5%/°C for EC
and from 2 to 7%/°C for resistivity
Probes (included) HI 76303
for conductivity measurements
HI 3316D
for resistivity measurements
Power Supply
12 Vdc (power adapter included)
Environment
0 to 50°C (32 to 122°F); RH max 95%
Dimensions
240 x 182 x 74 mm (9.4x7.1x2.9")
Weight
1.0 kg (2.3 lb.)
1) Liquid Crystal Display (LCD)
2) Conductivity calibration knob
3) Conductivity temperature coefficient knob
4) Range keys, to select the 199.9 μS/cm, 1999 μS/cm,
19.99 mS/cm or 199.9 mS/cm range
5) Resistivity temperature coefficient knob
6) ON/OFF switch
7) Power adapter socket
8) Probe connector
9) Resistivity calibration trimmer
FUNCTIONAL DESCRIPTION