Specifications
61
EC
TDS
Salinity
Temperature
compensation
Automatic -5 to 100 ºC (23 to 212 ºF)
NoTC - can be selected to measure
absolute conductivity.
Conductivity
temperature
coefficient
0.00 to 6.00% / ºC (for EC and TDS only). Default value is 1.90% / ºC
TDS factor
0.40 to 0.80 (default value is 0.50)
EC Probe
HI 763100
Log feature
Up to 1000 * (400) records organized in: Manual log on demand (Max. 200
logs), Manual log on stability (Max. 200 logs), Interval logging * (Max. 600
logs; 100 lots)
DO
Range
0.00 to 45.00 ppm (mg/L)
0.0 to 300.0%
0.0 to 50.0 ºC (32.0 to 122.0 °F)**
Resolution
0.01 ppm (mg/L)
0.1%
0.1 °C
Accuracy
@ 25 °C / 77 °F
±1.5% of reading or ±1 digit
±0.5 ºC
DO Calibration
One or two points at 0% (HI 7040) and 100% (water saturated
air)
Altitude Compensation
Resolution
-500 to 4,000 m (-1640 to 13120’)
100 m (328’)
Salinity Compensation
Resolution
0 to 40 g/L
1 g/L
Temperature
Compensation
0.0 to 50.0 ºC
(32.0 to 122 ºF)
DO Probe
HI 764080
Log Feature
Up to 1000 records organized in Manual Log on Demand
(Max. 200 logs)
Manual Log on Stability (Max. 200 logs)
Interval Logging 100 lots,
Max. 600 logs/lot
Additional Specifications
PC Interface
Micro USB
Storage Interface
USB
Power Supply
5 VDC Adapter (included)
Environment
0-50 ºC (32-122 ºF) Max 95% RH non-condensing
Dimensions
202 x 140 x 12 mm (7.9 x 5.5 z 0.5”)
Weight
250g (8.82 oz)
Specifications
*Standard Mode Only
** Temperature limits will be reduced to actual probe/sensor limits.
‡Absolute conductivity (or TDS) is the conductivity (or TDS) value without temperature
compensation.
Dissolved
Oxygen
Specifications
www.
.com
1.800.561.8187