Setup/Installation
18
Logged data on edge® can be transferred from the meter to a USB flash drive
by using the log recall function. The minimum requirement for the drive is USB
2.0. Select the pH, EC or DO record you wish to export and follow the simple
steps below.
Connect USB flash drive to the USB port, located on
the top of the meter. Press the
RCL
key and select
the parameter to view (pH, EC, or DO). Press the
CFM
key. Select Manual, Stability, or interval lots by using
the
pq
keys. Press the
LOG
key (not
CFM
). The “USB
HOST” tag should come on.
“PLEASE WAIT” message appears followed by
“EXPORT”. Press
CFM
to export the selected record or
lot. If
CFM
is not pressed in 10 seconds, the USB host
will become inactive.
The meter will display the percentage of export.
The export percentage should go to 100%. Remove
USB flash drive.
If the selected file is already saved on the flash drive,
edge® will ask for confirmation of overwriting the
existing file. The message “OVERWRITE” and “CFM”
tag will blink. Press
CFM
for overwriting the existing
file or
CAL
to exit without exporting.
After exporting the display will return to the
selected file. Press the
RCL
key twice to return to
measurements.
Note
: Do not remove USB flash drive during an active export transfer.