Operational Guide
34
EC Meter
Configuration
Option
Description
Choices
Default
Basic mode
Basic Mode
Off or On
Off
Available
Temperature
Compensation
The user
may select
Automatic
Temperature
Compensation
or No TC to
configure
absolute
conductivity.
No TC or ATC
ATC
Not available.
ATC is
automatically
used.
C.F (cm
-1
)
User may
enter actual
Cell factor
value Manual
calibration. (see
page 35)
0.01 to 9.999
cm
-1
1.000 cm
-1
Not available.
Automatically
determined
during
calibration.
T.Coef. (%/ºC)
This parameter
is related to the
solution being
measured at
temperatures
other than 20
or 25 ºC. It is
used to correct
measured
conductivity
to a reference
temperature by
applying a fixed
factor for linear
compensation.
0.00 to 6.00
(%/ºC) Note:
Setting to 0.00
is the same as
using No TC.
1.90 (%/ºC)
Close for natural
waters or salt
solutions.
Not available.
Automatically
set to 1.90%/ºC.
T.Ref. (ºC)
The user may
select either
20 ºC or 25 ºC
reference
temperature for
temperature
corrected
conductivity.
20 ºC or 25 ºC
25 ºC
Available
TDS Conversion
Factor.
This factor
is used to
mathematically
convert
conductivity to a
TDS value.
0.40 to 0.80
0.50
Not Available.
Automatically
set to 0.50.
View T.Ref. or
T.Coef.
The user may
choose whether
to display the
reference
temperature
(T.Ref) or the
Temperature
coefficient on
the display
with the
measurement.
T.Ref.(ºC) or
T.Coef.(%/ºC)
T.Ref (ºC)
T.Ref (ºC)
automatically
displayed during
measurement.