Setup/Installation

9

The  main  operating  modes  of  edge

DO 

®    are  setup,  calibration,  measurement, 

data  logging,  and  data  export.  Follow  this  general  outline  of  steps  to  get  you 
started. The following topics are expanded upon in the sections that follow in 
this manual.

1. 

Familiarize yourself with the design features of this unique meter.

2. 

Decide how the meter will be used and set up the wall or bench cradle in a 
clean area near line power.

3. 

Turn edge

DO 

® on using the ON/OFF button located on the top of the meter.

4. 

Plug in the probe required for measurement.

5. 

SETUP the measurement parameters required for the measurement you 
will be making.

6. 

Calibrate the sensor/probe.

You are now ready for measurements.

Bench Cradle Setup

Insert electrode holder arm into the post 
on the pivoting base.

Connect  the  probe  connector  to  the 
socket  located  at  the  bottom  of  the 
instrument.

Slide  edge

DO 

®  into  the  cradle  while 

positioning  the  probe  cable  behind  the 
cradle.  Put  the  probe/sensor  into  the 
electrode  holder  and  secure  cable  in 
clips.

Connect the power adapter cable to the 
rear socket of the bench cradle. Connect 
the other end to the power adapter and 
plug  into  line  power.  Verify  the  battery 
icon indicates charging.

Wall Cradle Setup

Choose  suitable  wall  location.  (Use  2.5 
mm or US #3 bit). Fasten the wall cradle 
using  the  provided  screws.  Snap  cover 
over screw heads.

Setup/
Installation

Setting Up 
edge

DO 

®

Summary of Contents for edge DO

Page 1: ...DO...

Page 2: ...of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Included 3 Safety Measures 3 Description 4 Diagram 4 8 Product Diagram 4 5 Probe Diagram 6 Keypad Function 7 Guide to Ind...

Page 3: ...rings for membrane caps Before using this product make sure that it is entirely suitable for your specific application and for the environment in which it is used Operation of this instrument may caus...

Page 4: ...libration measurement data logging and transfer of data to a USB thumb drive or computer Every feature and measurement detail is designed to give you an edge in measurement technology edgeDO is versat...

Page 5: ...2 Capacitive Touch Keypad 3 3 mm jack input for edgeDO digital probes 4 Top mounted ON OFF button 5 Micro USB device connection for power or PC interface 6 Standard USB host connection for data transf...

Page 6: ...sensor recognition Store calibration specific data from the last calibration Are built with materials suitable for use in chemical analysis Have integrated temperature measurement Incorporate a 3 mm...

Page 7: ...TUP CLR Used to enter exit SETUP mode During calibration used to clear previous calibration data In log RCL used to clear log records 5 Used to scroll through SETUP menu Used to change selection when...

Page 8: ...nt 10 Temperature units 11 Temp Compensation status 12 Measurement line 13 Measurement units 14 Stability Indicator The third line of the LCD 9 is a dedicated message line During measurement the user...

Page 9: ...d for the measurement you will be making 6 Calibrate the sensor probe You are now ready for measurements Bench Cradle Setup Insert electrode holder arm into the post on the pivoting base Connect the p...

Page 10: ...t the top Plug the 5 VDC adapter into the power supply socket or by connecting directly to a PC Note edgeDO is supplied with a rechargeable battery inside which provides about 8 hours of continuous us...

Page 11: ...ata will be exported to the PC LOG ON EDGE or EXPORT TO PC LOG ON EDGE Log Select log type to be used from 3 types of logging Manual log on demand Manual log on stability 3 types of stability criteria...

Page 12: ...lash Format Only seen when log errors are present Permits formatting the flash drive On or Off OFF Message Transition User may choose how messages are displayed on third LCD line of display Word scrol...

Page 13: ...urement type New records made on different days are stored in the same Manual lot Manual Stability log on demand A log on demand that is made each time LOG key is pressed and the stability criteria is...

Page 14: ...top the Interval logging session The LOG STOPPED message will be displayed for a few seconds If a sensor failure occurs during interval logging the message OUT OF SPEC will alternate with logging info...

Page 15: ...pressing LOG again When the stability selected criteria has been met a SAVED message will be displayed followed by a screen indicating how much log space is available The LOG tag will be displayed on...

Page 16: ...te Press CLR The instrument will display CLEAR MANUAL for Manual Records CLEAR STAB for Stability Records For Interval lots the message CLEAR followed by the selected lot will be displayed with CFM ta...

Page 17: ...were deleted The total number of lots that can be saved is 100 If some are deleted for example 1 50 fifty additional logs may be stored These will be numbered 101 150 The lots are allocated successiv...

Page 18: ...followed by EXPORT Press CFM to export the selected record or lot If CFM is not pressed in 10 seconds the USB host will become inactive The meter will display the percentage of export The export perc...

Page 19: ...ote Western Europe ISO 88859 1 character set and English language are suggested settings Other files may be visible depending upon computer settings All files stored will appear in this folder Adjust...

Page 20: ...the cardboard shipping tube used to protect the probe during shipping Save the tube should the probe be stored dry again 2 Open membrane package and remove one O ring and one membrane cap 3 Rinse the...

Page 21: ...reaches an equilibrium with the surrounding solution Whenever measurements are taken with a non polarized probe the measurement will be drifty and inaccurate The measurement will jump when the probe...

Page 22: ...the most accurate calibration and concentration measurement 500 400 300 200 100 0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 2...

Page 23: ...18 less compared to fresh water at 25 C By entering the approximate salinity value the calibration and subsequent concentration measurement will be compensated to display the correct oxygen concentrat...

Page 24: ...leave calibration after the first point The instrument will display SAVING message and it will return to measurement mode memorizing the slope calibration data Zero Calibration PressCALorcontinuewith...

Page 25: ...standard Altitude and Salinity factors time and date of the last calibration the expired calibration information and the probe serial number This information is also included with all logged data Last...

Page 26: ...2 DAYS AGO The probe serial number together with the current reading Make sure that the probe is polarized calibrated and the protective cap has been removed Rinse probe Submerse the probe in the samp...

Page 27: ...e between the temperature at which the probe was stored and the temperature of the sample the longer the time will be If the DO temperature exceeds the limits of the probe the message PROBE OUT OF SPE...

Page 28: ...ow The platinum cathode DO Probe Diagram page 21 should always be bright and untarnished If it is tarnished or stained the cathode should be cleaned You can use a clean lint free cardboard or cloth Ru...

Page 29: ...t and clean if necessary Install new cap fresh electrolyte with no bubbles permit longer polarization Stir or increase flow rate The meter does not measure temperature is displayed on second LCD line...

Page 30: ...g L 1 g L Temperature Compensation 0 0 to 50 0 C 32 0 to 122 F DO Probe HI 764080 Log Feature Up to 1000 records organized in Manual Log on Demand Max 200 logs Manual Log on Stability Max 200 logs Int...

Page 31: ...30 mL HI 764080 Spare DO probe HI 764080A P 5 spare membranes Other Accessories HI 75110 220U Voltage adapter from 115 VAC to 5 VDC USA plug HI 75110 220E Voltage adapter from 230 VAC to 5 VDC Europe...

Page 32: ...l Hanna Office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges i...

Reviews: