Setup/Installation
19
Logged data on the edge
DO
® can be transferred from the meter to a PC by
following these simple directions. Suitable operating systems include Windows
(Xp minimum), OS X or Linux.
1.
Connect edge
DO
® to the PC using the supplied micro USB cable.
2.
Turn on edge
DO
®.
3.
Press
SETUP
and select “LOG ON EDGE”.
4.
Press
MODIFY
then use
pq
keys to change to “EXPORT TO USB”.
5.
Press
CFM
and the USB/PC Tag is displayed.
6.
Press
SETUP
to exit.
The PC should detect the USB as a removable drive. Open the drive to view the
stored fi les. Log fi les are formatted as Comma separated values (*.CSV) and can
be opened with any text editor or spreadsheet program.
Note
:
•
Western Europe (ISO-88859-1) character set and English language are
suggested settings.
•
Other fi les may be visible depending upon computer settings. All fi les
stored will appear in this folder.
•
Adjust Font or column width appropriately.
Interval logs are designated as DO Lots. ie. DOLOT001, DOLOT002, DOLOT003.
The Manual Lots are DOLOTMAN.
The Stability Lots are DOLOTSTAB. All stability logs, regardless of stability
setting, are located in the same stability fi le for that measurement.
Click on the desired log to view data.
Note
:
•
If “ºC !” appeared in log data, the electrode/probe was used beyond it’s
operation specifi cations and the data is not considered reliable.
•
If “ºC !!” appeared in log data, the temperature sensor within the probe or
electrode is broken and the device should be replaced. Logged data should
not be considered reliable.
PC & Storage
Interface