Tips for an Accurate Measurement

• Ensure the sample does not contain any debris.

• Hold dropper bottles completely vertically and squeeze slowly 

to ensure correct drop size.

• Whenever the cuvette is placed into the checker, it must be dry 

outside and free of fingerprints, oil and dirt. 

• Wipe the cuvette thoroughly with 

HI731318

 microfiber 

cleaning cloth or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher 

readings. To obtain accurate measurements, remove such 

bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent has 

been added, as accuracy will be affected.

• Discard the sample immediately after the reading has been 

taken or the glass might become permanently stained.

• Prepare sample cuvette immediately after collecting sample from tank. Ammonia 

is volatile and will dissipate if stored in a bottle before analysis, causing low 

measurements.

Calculating Unionized Toxic Ammonia

This method measures NH 

3

 / NH 

4

+

. To determine the toxic NH 

3

 portion, use the table 

to determine the percentage of toxic NH 

3

 . Levels of toxic NH 

3

 greater than 0.01 ppm 

have a negative effect on fish. Multiply total ammonia (NH 

3

 / NH 

4

+

) by the percent 

unionized ammonia from the table to determine unionized toxic ammonia. 

Example

: pH 8.0, Temperature: 24 °C (75 °F), NH 

3

 / NH 

4

+

 1.00 ppm

Unionized toxic ammonia: 1.00 ppm x (5.0 % ÷ 100 )= 0.05 ppm

Percentage of Unionized Ammonia (

NH 

3

)

pH

21 °C (70 °F)

24 °C (75 °F)

26 °C (79 °F)

29 °C (84 °F)

7.6

1.7

2.1

2.4

2.9

7.8

2.6

3.2

3.7

4.5

8.0

4.1

5.0

5.8

7.0

8.2

6.3

7.7

8.8

11

8.4

9.7

12

13

16

8.6

15

17

20

23

8.8

21

25

28

32

10mL

10mL

Battery Replacement

To save the battery, the checker shuts down after 10 minutes 

of non-use.

A fresh battery lasts for a minimum of 5000 measurements. 

When the battery is drained, the instrument displays “bAd” 

then ”bAt”, and turns off. 
To replace the battery, follow the next steps:
1. Press and hold the ON/OFF button to turn the checker off.
2. Turn the instrument upside down and use a screwdriver to 

unfasten the screw and remove the battery cover.

3. Remove the old battery, replace it with a new 1.5V AAA 

battery, inserting the negative end first.

4. Replace the battery cover, fasten and tighten the screw.

Accessories

Reagent Sets

HI784-25

Reagents for 25 Marine Ammonia tests

Other Accessories

HI784-11

Marine Ammonia certified standard kit

HI731315

Glass cuvette and cap for Checker

®

HC colorimeters (2 pcs.)

HI731318

Cloth for wiping cuvettes (4 pcs.)

HI740028P

1.5V AAA battery set (12 pcs.)

HI740157P

Transfer pipette (20 pcs.)

HI93703-50

Cuvette cleaning solution, 230 mL

Certification

All Hanna

®

 instruments conform to the 

CE European Directives.

Disposal of Electrical & Electronic Equipment.

 The product 

should not be treated as household waste. Instead, hand it over to 
the appropriate collection point for the recycling of electrical and 
electronic equipment, which will conserve natural resources.

HI784

Marine Ammonia (NH 

3

 / NH 

4

+

)

INSTRUCTION MANUAL

Disposal of waste batteries.

 This product contains batteries, do not dispose of 

them with other household waste. Hand them over to the appropriate collection 
point for recycling.
Ensuring proper product and battery disposal prevents potential negative 
consequences for the environment and human health. For more information, contact 
your city, your local household waste disposal service, or the place of purchase.

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific 
application and for the environment in which it is used. Any variation introduced 
by the user to the supplied equipment may degrade the checker’s performance. 
For your and the checker’s safety do not use or store it in hazardous environments.

Warranty

HI784

 Checker

 

HC is warranted for a period of one year against defects in 

workmanship and materials when used for its intended purpose and maintained 
according to instructions. This warranty is limited to repair or replacement 
free of charge. Damage due to accidents, misuse, tampering, or lack of 
prescribed maintenance is not covered. If service is required, contact your local 
Hanna Instruments

®

 office. If under warranty, report the model number, date of 

purchase, serial number, and the nature of the problem. If the repair is not covered 
by the warranty, you will be notified of the charges incurred. If the checker is to be 
returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) 
number from the Technical Service department and then send it with shipping 
costs prepaid. When shipping any product, make sure it is properly packaged for 
complete protection.

Hanna Instruments reserves the right to modify the design, construction, or 
appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the 

written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, 

Rhode Island, 02895, USA.

IST784 02/22

Summary of Contents for Checker Marine HI784

Page 1: ...HI784 Marine Ammonia NH3 NH4 INSTRUCTION MANUAL ...

Page 2: ...port team Each HI784 is delivered in a case with custom insert and is supplied with Marine Ammonia reagent starter kit reagents for 10 tests Sample cuvette and cap 2 pcs Plastic refilling pipette 1 pc 1 5V AAA Alkaline battery 1 pc Instruction manual Quick reference guide Note Save all packing material until you are sure that the Checker HC handheld colorimeter works correctly Any damaged or defec...

Page 3: ...t causes a blue green tint in the sample Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Prepared sample cuvette sample plus reagents must be 18 to 29 C 65 to 85 F Battery type 1 5V AAA Alkaline Auto shut off After 10 minutes of non use Dimensions 86 0 x 61 0 x 37 5 mm 3 4 x 2 4 x 1 5 Weight 64 g 2 3 oz Warm or cool prepared cuvettes if needed Functional Description LCD Display Cap Cuve...

Page 4: ...e and invert for 30 45 seconds to mix Do NOT shake vigorously this will cause more air bubbles For the most accurate reading ensure all reagent is dissolved and there are no visible bubbles Ensure the outside of the cuvette is dry and clean Let the cuvette stand for 30 seconds This allows complete dissipation of micro bubbles Insert the cuvette into the checker and close the cap Press the ON OFF b...

Page 5: ...des an explanation of the errors and warnings and the recommended action to be taken Light High There is an excess amount of ambient light reaching the detector Please check preparation of zero cuvette Light Low There is not enough light to perform a measurement Please check the preparation of the zero cuvette Under Range Minimum concentration value displayed blinking indicates the measured value ...

Page 6: ...ng has been taken or the glass might become permanently stained Preparesamplecuvetteimmediately after collecting sample from tank Ammonia is volatile and will dissipate if stored in a bottle before analysis causing low measurements Calculating Unionized Toxic Ammonia This method measures NH3 NH4 To determine the toxic NH3 portion use the table to determine the percentage of toxic NH3 Levels of tox...

Page 7: ...4 Replace the battery cover fasten and tighten the screw Accessories Reagent Sets HI784 25 Reagents for 25 Marine Ammonia tests Other Accessories HI784 11 Marine Ammonia certified standard kit HI731315 Glass cuvette and cap for Checker HC colorimeters 2 pcs HI731318 Cloth for wiping cuvettes 4 pcs HI740028P 1 5V AAA battery set 12 pcs HI740157P Transfer pipette 20 pcs HI93703 50 Cuvette cleaning s...

Page 8: ...its intended purpose and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not covered If service is required contact your local Hanna Instruments office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair...

Reviews: