Measurement Procedure

1. Press the ON/OFF button to turn the checker on. 

All segments will be displayed for a few seconds, 
followed by “Add”

“C.1” with “Press” blinking.

2. Preparing the cuvette for “C.1” measurements:
• Use the pipette to fill the cuvette with 10 mL of 

unreacted sample.

• Hold the 

HI784A-0

 Reagent A dropper bottle 

vertically and add 14 drops to the cuvette.

Note: 

For best results, tap the dropper on a hard surface 

and clean the outside of the tip prior to use.

• Using scissors, open one packet of 

HI784B-0

 

Reagent B along the dotted line. Push the two corners 
together to make a spout.  Add the content of the 
packet to the cuvette.

• Screw the cap onto the cuvette and invert for 

30-45 seconds to mix. Do NOT shake vigorously – 
this will cause more air bubbles! 
For the most accurate reading, ensure all reagent is 
dissolved and there are no visible bubbles. Ensure the 
outside of the cuvette is dry and clean.

• Let the cuvette stand for 30 seconds. This allows 

complete dissipation of micro-bubbles.

• Insert the cuvette into the checker and close the cap.
• Press the ON/OFF button. When the display shows 

“Add”, “C.2” with “Press” blinking, the checker is 
zeroed. Remove the cuvette.

3. Preparing the cuvette for “C.2” measurements:
• Unscrew the cuvette cap. Hold the 

HI784C-0

 

Reagent C dropper bottle vertically and add 12 drops 
to the cuvette.

• Replace the cap and invert the cuvette 5 times to mix. 
• Insert the cuvette into the checker and close the cap.

B

10mL

10mL

C.1

8:.8.8

10mL

10mL

10mL

10mL

10mL

10mL

C.2

10mL

10mL

10mL

10mL

10mL

10mL

10mL

10mL

Dear Customer, 

Thank you for choosing a Hanna Instruments

®

 product. Please read this 

instruction manual carefully before using the Checker

®

HC handheld 

colorimeter. For more information about Hanna Instruments and our 
products, visit www.hannainst.com or e-mail us at [email protected].
For technical support, contact your local Hanna Instruments office or e-mail 
us at [email protected].

Preliminary Examination

Remove the Checker

 

HC handheld colorimeter and accessories from the packing 

material and examine it carefully. If you require any further information, please 
contact Hanna Instruments technical support team.
Each 

HI784

 is delivered in a case with custom insert and is supplied with:

• Marine Ammonia reagent starter kit (reagents for 10 tests)
• Sample cuvette and cap (2 pcs.)
• Plastic refilling pipette (1 pc.)
• 1.5V AAA Alkaline battery (1 pc.)
• Instruction manual
• Quick-reference guide

Note: 

Save all packing material until you are sure that the Checker

 

HC handheld 

colorimeter works correctly. Any damaged or defective item must be returned in its 

original packing material with the supplied accessories.

General Description & Intended Use

HI784

 Marine Ammonia handheld checker is designed to determine the 

concentration of ammonia in saltwater aquariums. 

HI784

 features a single-button operation system and is easy to use.

The large LCD is easy to read and the auto shut-off feature assures the battery 
will not be drained.

Specifications

Range

0.00 to 2.50 ppm (mg/L) NH

3

Resolution

0.01 ppm (mg/L)

Accuracy

±0.05 ppm ±5% of reading @ 25 °C (77 °F)

Light source

Light Emitting Diode @ 610 nm

Light detector

Silicon photocell

Method

Adaptation of the Salicylate Method. The reaction between 
Ammonia and Ammonium and the reagent causes a 
blue-green tint in the sample. 

Environment

0 to 50 °C (32 to 122 °F); max. 95% RH non-condensing
Prepared sample cuvette (sample plus reagents) must be 
18 to 29 °C (65 to 85 °F)*.

Battery type

1.5V AAA Alkaline

Auto shut-off

After 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

64 g (2.3 oz)

* Warm or cool prepared cuvettes if needed.

Functional Description & LCD Display

Cap
Cuvette with cap

Cuvette holder

LCD (Liquid Crystal Display)

ON/OFF button

Ammonia

ppm NH

3

/NH

4

+

• Press and hold the ON/OFF button. The display will show the countdown prior 

to the measurement. Alternatively, wait 15 minutes and press the button once 
to measure.

 

0:15

 

1.05

Note: 

For samples with less than 1.00 ppm, the reaction is complete after 10 

minutes.  Users may press the ON/OFF button at this time, with 5 minutes 

remaining on the timer.

• When the timer ends, the checker will perform the reading. The instrument 

displays the ammonia/ammonium concentration in ppm NH

3

. The checker 

automatically turns off 10 minutes after reading.

• To calculate the unionized ammonia (NH

3

) concentration, use the conversion 

table on the reverse side.

Errors & Warnings

The checker shows clear warning messages when erroneous conditions appear and 
when measured values are outside the expected range. The information below 
provides an explanation of the errors and warnings, and the recommended action 
to be taken.

Light High:

 There is an excess amount of ambient light 

reaching the detector. Please check preparation of zero cuvette.

Light Low:

 There is not enough light to perform a 

measurement. Please check the preparation of the zero cuvette.

Under Range:

 Minimum concentration value displayed 

blinking indicates the measured value is outside the limits of the 

method. Verify that the sample does not contain any debris, and 

the preparation of the sample cuvette.

Over Range:

 Maximum concentration value displayed blinking 

indicates the measured value is outside the limits of the method. 

Verify the preparation of the sample cuvette. Dilute the sample 

and repeat the measurement.

Battery Low:

 Battery level is too low for the checker to function 

properly. Replace the battery with a new one.

Drained Battery:

 The battery is drained and 

must be replaced. Replace the battery with 
a new one and restart the checker.

L.Hi

L.Lo

0.00

2.50

bAt

bAd bAt

Summary of Contents for Checker Marine HI784

Page 1: ...HI784 Marine Ammonia NH3 NH4 INSTRUCTION MANUAL ...

Page 2: ...port team Each HI784 is delivered in a case with custom insert and is supplied with Marine Ammonia reagent starter kit reagents for 10 tests Sample cuvette and cap 2 pcs Plastic refilling pipette 1 pc 1 5V AAA Alkaline battery 1 pc Instruction manual Quick reference guide Note Save all packing material until you are sure that the Checker HC handheld colorimeter works correctly Any damaged or defec...

Page 3: ...t causes a blue green tint in the sample Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Prepared sample cuvette sample plus reagents must be 18 to 29 C 65 to 85 F Battery type 1 5V AAA Alkaline Auto shut off After 10 minutes of non use Dimensions 86 0 x 61 0 x 37 5 mm 3 4 x 2 4 x 1 5 Weight 64 g 2 3 oz Warm or cool prepared cuvettes if needed Functional Description LCD Display Cap Cuve...

Page 4: ...e and invert for 30 45 seconds to mix Do NOT shake vigorously this will cause more air bubbles For the most accurate reading ensure all reagent is dissolved and there are no visible bubbles Ensure the outside of the cuvette is dry and clean Let the cuvette stand for 30 seconds This allows complete dissipation of micro bubbles Insert the cuvette into the checker and close the cap Press the ON OFF b...

Page 5: ...des an explanation of the errors and warnings and the recommended action to be taken Light High There is an excess amount of ambient light reaching the detector Please check preparation of zero cuvette Light Low There is not enough light to perform a measurement Please check the preparation of the zero cuvette Under Range Minimum concentration value displayed blinking indicates the measured value ...

Page 6: ...ng has been taken or the glass might become permanently stained Preparesamplecuvetteimmediately after collecting sample from tank Ammonia is volatile and will dissipate if stored in a bottle before analysis causing low measurements Calculating Unionized Toxic Ammonia This method measures NH3 NH4 To determine the toxic NH3 portion use the table to determine the percentage of toxic NH3 Levels of tox...

Page 7: ...4 Replace the battery cover fasten and tighten the screw Accessories Reagent Sets HI784 25 Reagents for 25 Marine Ammonia tests Other Accessories HI784 11 Marine Ammonia certified standard kit HI731315 Glass cuvette and cap for Checker HC colorimeters 2 pcs HI731318 Cloth for wiping cuvettes 4 pcs HI740028P 1 5V AAA battery set 12 pcs HI740157P Transfer pipette 20 pcs HI93703 50 Cuvette cleaning s...

Page 8: ...its intended purpose and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not covered If service is required contact your local Hanna Instruments office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair...

Reviews: