1.3. Precautions for safety
T04
Centrifugal separators are dangerous because they use a high-speed rotating body.
Safety precautions are to prevent personal injury, product damage,
and breakdown from possible dangers during use.
Please observe all safety measures described in this manual.
1. The centrifuge must be installed horizontally on a level surface. If the shaft is operated in an
inclined state, vibration or damage to the device may occur.
2. Before connecting the centrifuge to power, check the voltage to be used. If it is connected
with the wrong voltage, it may cause damage to the device and personal injury.
3. Use only the rotor and recommended parts and accessories provided by Hanil Scientific Inc.
We are not responsible for any damage to the device or accidents resulting from the use of
non- recommended parts and accessories.
4. The sample must be used in a centrifugal separator tube, and must be used within the
maximum xg value of the tube.
5. In the case of centrifuging dangerous substances (pathogenic, toxic, radioactive substances,
etc.), it is necessary to sufficiently grasp the physical properties of the substance and take
necessary safety measures.
6. If the centrifuge is contaminated with pathogenic, toxic or radioactive substances, the
contaminants must be thoroughly removed and necessary measures such as ventilation or
isolation must be considered.
7. Substances that can generate volatile or explosive vapors cannot be centrifuged.
8. When the rotor comes into contact with cleaning solutions such as strong acids or strong
bases or cesium/silver/salt, it will cause a chemical reaction and corrosion will begin.
9. The rotor chamber must be kept dry at all times before using the centrifuge.
10. Do not exceed the maximum RPM / RCF. If the rotor is subjected to a centrifugal force that
exceeds the allowable rotation limit, the rotor will be deformed and damaged.
11. Before centrifugation, the sample must be balanced.
12. It is forbidden to touch or move the rotating rotor.
13. The rotor must be accurately fixed to the rotating shaft, and the rotor used with the rotor lid
must be securely fastened. If the lid comes off during rotation, it may cause serious
damage to the Centrifuge, sample, and product.
14. Do not block the air ventilation for proper air flow which keep the centrifuge from overheating.
15. Do not put any objects into the openings of the centrifuge.
16. Never use a tool to remove the lid or guard.
17. When requesting repair, the user must remove contaminants in advance.
18. Maintenance must be performed by a technician authorized by Hanil Scientific Inc.
19. For product repair, contact the place of purchase.
20. When operating according to the IEC61010-2-020 standard, the safety distance (30 cm)
around the centrifuge must be observed for smooth instrument operation and the safety of
users and the surrounding environment.
21. Turn off the device switch after using the device.
22. Disconnect and store the power cord from the power outlet before cleaning the machine or
when not in use for a long time.
T. +82-2-3472-0727 │ [email protected] │ www.ihanil.com
Summary of Contents for T04
Page 1: ...T04 LOW SPEED CENTRIFUGE USER MANUAL...
Page 7: ...ihanil com 6 USER MANUAL T04 T 82 2 3472 0727 sales ihanil com www ihanil com...
Page 8: ...7 ihanil com USER MANUAL T04 T 82 2 3472 0727 sales ihanil com www ihanil com...
Page 23: ...MEMO ihanil com 22 USER MANUAL T04 T 82 2 3472 0727 sales ihanil com www ihanil com...
Page 24: ......