Hanil T04 User Manual Download Page 11

USER MANUAL 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

       

 

 

 

 

 

     

   

 

 

 

 

 

 

 

   

 

   

 

 

 

 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

   

 

 

 

   

 

 

 

 

   

 

   

 

   

 

   

 

 

 

 

 

 

   

 

   

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

   

 

 

   

   

 

 

 

 

 

   

 

   

   

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

   

 

 

 

 

 

 

 

 

 

     

 

 

 

   

 

 

 

 

 

 

 

 

 

 

   

 

 

   

 

 

 

 

   

 

   

 

 

   

 

 

   

   

   

 

 

   

   

ihanil.com 10

Before use, make sure that the rotor is firmly connected to the motor shaft.

3.5 Tube mounting

T04

3. Mount the 15ml stainless steel sleeve into all holes

of the rotor.

Before operation, make sure to mount all the sleeves

on the rotor.

1. Before inserting the sample tube, check that there is no foreign matter or moisture inside

the rotor hole or bucket.

If there is any foreign matter or moisture, be sure to remove it with a dry cloth.

Load samples in the tubes before placing the tubes into the rotor holes.

· Use the tube for centrifuge recommended by Hanil Scientific Inc, and do not use it
above the allowable standard after checking the maximum RCF value for each tube.

·Samples should be accurately measured in the same amount/density and placed in
each tube, then the tubes must be mounted on the rotor so that they are symmetrical
to each other. At this time, if the volume of the sample to be symmetrical is different,
strong vibration may occur when the rotor rotates or serious damage to the rotor and
motor may occur.

·Weight balance should be performed to minimize the difference in weight of the
tubes containing the sample.

·Even if the number of tubes is the same, they will be asymmetrical according to their
positions, so when inserting tubes into centrifugal tubes facing each other, you must
also check the positions. If the number of tubes is not symmetrical, you must use an
extra tube and insert it to balance the same weight as the other tubes.

T. +82-2-3472-0727 │ [email protected] │ www.ihanil.com

Summary of Contents for T04

Page 1: ...T04 LOW SPEED CENTRIFUGE USER MANUAL...

Page 2: ...Research Use Only www ihanil com LOW SPEED CENTRIFUGE T04 Product name Centrifuge Model name T04 Manufacturer Hanil Scientific Inc Address 16 Arayukro Gimpo 10136 Korea UM T04 E Rev 0 20210121...

Page 3: ...ting 10 1 Safety warnings and cautions 3 1 1 Safety labels 3 1 2 Conditions of use 3 1 3 Precautions for safety 4 4 How to use 12 4 1 Control Panel 12 4 2 Speed setting 13 4 3 Time setting 14 4 4 Star...

Page 4: ...sed product Please read carefully before using the product Rotor insertion tightning caution indication warning In case of Emergency the hole position to open the lid manually Lid closure caution indi...

Page 5: ...apors cannot be centrifuged 8 When the rotor comes into contact with cleaning solutions such as strong acids or strong bases or cesium silver salt it will cause a chemical reaction and corrosion will...

Page 6: ...ol Emergency Lid Open Tool 2 Product composition and information Grease Lubricant 2 1 Appearance 2 2 Components T04 Display Control Power Swich Power Socket Fuse Lid T04 is provided with AO 15 10 12 R...

Page 7: ...ihanil com 6 USER MANUAL T04 T 82 2 3472 0727 sales ihanil com www ihanil com...

Page 8: ...7 ihanil com USER MANUAL T04 T 82 2 3472 0727 sales ihanil com www ihanil com...

Page 9: ...n will be displayed with a beep sound When the device is shipped the default value is Max rpm 4 000 10 minutes 3 2 Power connection 1 Connect the AC power cord to the power socket located on the back...

Page 10: ...side the chamber holding the rotor stable when released The buffer must be removed before operating the Centrifuge Used to open the lid When the Lid is closed the lamp is off and when it is open the l...

Page 11: ...ximum RCF value for each tube Samples should be accurately measured in the same amount density and placed in each tube then the tubes must be mounted on the rotor so that they are symmetrical to each...

Page 12: ...ally The tube filled with the sample should have no difference in weight and the density should be symmetrically and evenly arranged Tubes for centrifuges must not exceed Max RCF value It should be us...

Page 13: ...et in minute and second and up to 99 minutes 59 seconds or continuous operation is possible Control interval 1 Min 1 Sec START STOP It is used to start and stop centrifugation Control interval 1 PROG...

Page 14: ...seconds will change in units of 100rpm 10xg If you do not enter the setting value for 15 seconds the setting mode is canceled 4 2 Speed setting T04 The speed setting value is displayed as RPM and RCF...

Page 15: ...ssing Time Once Twice Minuite Second input mode Entering the setting mode Min Sec lamp will turn on 2 Enter the setting value using the key and press Time Min Sec can be changed 1min 1sec Pressing the...

Page 16: ...If you want to end the operation press the Start Stop button During deceleration Start Stop lamp is turned off End of operation Start Stop lamp is turned off Final stop When operation is stopped the...

Page 17: ...diately as needed 1 After setting the program that needs to be saved press the PROG button for more than 3 seconds Display Save flashing 2 Using the key select the program number and press PROG again...

Page 18: ...is designed to call the program stored between 00 and 10 1 Press PROG once Display CALL flashing 4 7 Manual lid opening in case of emergency This method is used to remove the sample mounted on the rot...

Page 19: ...a soft cloth and wipe it with a dry cloth to prevent moisture from remaining 2 Do not use chemicals such as alcohol benzene benzol as they may damage it 3 Be careful not to scratch the surface while c...

Page 20: ...Rotor Mounting and Removal to install the rotor correctly If the tube insertion is asymmetrical or the weight is not correct refer to 3 5 Mounting the sample tube to check the tube weight and insert i...

Page 21: ...USER MANUAL T04 6 2 Error message information If the problem persists after taking the following measures please contact the service center ihanil com 20 1 Check if the lid closed 2 Check the Lid lam...

Page 22: ...21 ihanil com USER MANUAL T04 7 Rotor Accessories T 82 2 3472 0727 sales ihanil com www ihanil com...

Page 23: ...MEMO ihanil com 22 USER MANUAL T04 T 82 2 3472 0727 sales ihanil com www ihanil com...