3.1 Packing Inspection
3. Installation
• Check packing conditions carefully, before unpacking.
• Contact Hanil Scientific INC. immediately if damages found
• Check the delivery for completeness.
• You can get contact details on packing boxes and the back of the manual.
3.2 Installation
Installation on hard and flat ground.
•
Centrifuge should be installed on hard and flat place.
•
If the centrifuge is installed in an inclined place, the shaft may be bent due to
the weight of the rotor.
Good ventailation.
•
For air circulation and safety, maintain a 30cm “clearance envelope” around
the centrifuge while the rotor is spinning.
Constant temperature/
humidity
•
Centrifuge equipped with the sensitive electronic software which is fragile with
humidity and temperature.
•
Must avoid direct ray or heater and be put in the ambience of controlled temp.
and air
Avoid the corrosive gas
•
Install the centrifuge in a place where corrosive gas is not generated.
•
Sulfur dioxide gas and chlorine gas may cause corrosion
Leveling
•
The shaft should be put exactly vertical on the hotizontaliy flat ground by
the leveling tool
3.2.1 Selecting the location