background image

Operating Manual

Smart 15 Plus

Micro Centrifuge

Copyright 

Hanil Scientific INC. All right reserved.

Summary of Contents for Smart 15 Plus

Page 1: ...Operating Manual Smart 15 Plus Micro Centrifuge Copyright ⓒ Hanil Scientific INC All right reserved ...

Page 2: ...tion panel 8 4 2 Opening and closing the lid 9 4 3 Loading and removal of the rotor 9 4 4 Setting the Rotor ID 9 4 5 Loading tubes 10 4 6 Setting RPM RCF 11 4 7 Setting run time 11 4 8 Setting Accelation Deceleration rate 11 4 9 Starting the centrifugation run 11 4 10 End the centrifugation 11 4 11 Pulse Funcion 12 4 12 Saving Recall a program 12 5 Maintenance 13 5 1 Care instructions 13 5 2 Clean...

Page 3: ...dspeedoftherotororbucketsinuse 7 Substancesthatmaygeneratevolatileorexplosivevaporcannotbe centrifuged 8 Thebalancingworkofsamplesshoudbedoneinadvancebeforeoperation 6 Ifthecentrifugeiscontaminatedbytoxicorradioactivesamplesorinfectiousbloodsamples remove contaminantscompletely and take needful actions such as ventilationor isolation of centrifuge 9 Beforeoperation rotorandchambershouldbedry 10 Do...

Page 4: ...icating a place in danger of electric shock 1 4 Electric safety information 2 Usesockets withaprotective earthconductor andsuitablepower cord 1 Useapower cord onlyprovidedwithequipment 3 Donotuse anextension cord 4 Donotplaceanything onthe power cable 5 Donotblockvents 6 Ifyou havethefollowingemergencies shut offthepower supply andunplug thepower cordfrom outlet and contact your placeof purchase U...

Page 5: ...k Chamber Where the rotor is loaded Lid packing Rotor Connected to the motor shaft for centrifugalseparation Operation panel Display and keys for operating the centrifuge Power switch Switch for switching the centrifuge on and off Emergency Lid Open system 2 2 Delivery package Smart R15 Plus Main body Power cable User manual Rotor Locking Tool T wrench Extra fuse www ihanil com 4 ...

Page 6: ...Max RPM Max RCF Max Capacity Time DEC ramps Power supply Dimension W x D x H mm Weight kg Cat No ACC ramps 24 x 1 5 2 0ml 21 055 xg 99min 59sec 6 steps 6 steps AC 220V 50Hz 60Hz 293 x 380 x 242 mm 15 Without rotor SM 15PL techsupport ihanil com 5 ...

Page 7: ...tor Good ventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidity Centrifuge equipped with the sensitive electronic software which is fragile with humidity and temperature Must avoid direct ray or heater and be put in the ambience of controlled temp and air Avoid the corrosive gas Install the cent...

Page 8: ...ndto theelectricalrequirementsonthelabel attachedtothedevice Usesocketswithaprotectiveearthconductorandprovidedpowercord 1 Connectthe powercabletothepowersocketon therightsideoftheproductandplugthecordintothe outlet 2 Switchthecentrifugeon usingthepowerswitchontherightsideofthedevice techsupport ihanil com 7 ...

Page 9: ...l Save Door Start Stop Short Set upinformation operating conditionisdisplayed Setthe speedor runtime Moves thecursorleftorright Save the setting value Save or load the programs Release the lid Starting and stopping or short run centrifugation Setthe speedor runtime www ihanil com 8 ...

Page 10: ...n the motor shaft and the bore of the rotor Load the rotor vertically onto the morot shaft Tighten the clamping of the rotor with wrench by turning clockwise Check the rotor to make sure it loaded firmly Loading Loosen the clamping of the rotor by turning counter clockwise with the wrench Lift up the rotor from the motor shaft 4 4 Setting the Rotor ID Beforetheoperation checkandsettherotorID Press...

Page 11: ...ldbeloaded symmetrically Donotexceedthemaximumratedspeedofthe tube Samevolumeofsampleshouldbeputon tubes Checksymmetricloadingbybalncingtubeswithscales Check the maximum load for each tube Put tubes into rotor holes Tubes located opposite each other must be the same type and contain the same quantity www ihanil com 10 ...

Page 12: ...r Down key to set the value Press the Setup Enter key to save the settings 4 8 Setting Accelation Deceleration rate Press the Setup Enter key Accel Decel step is 0 5 Step 1 is the slowest and step 5 is the fastest Decel step 0 means natural brake 4 9 Starting the centrifugation run Check the all parameters are correct and the lid is closed 4 10 End the centrifugation Press Start Stop key to abort ...

Page 13: ... pressing Up down key 4 Press the Setup Enter key to save the program If you release the Start Stop key it stops when set RPM is reached 4 12 Saving Recallprograms 1 Set the centrifugationtime speed Accel Decel values You can store 10 programs on the device 0 9 Recall a programs 1 Press Call Save key 2 Press Up down key to select the program slot you want 3 Press the Setup Enter key to call up a p...

Page 14: ...nol Use the stopwatch to check that the time setting is correct If the device is contaminated use a mild cleaning fluid to clean Make sure do not scratch the surface of equipment when cleaning it Do not use a metal sponge If the device is rusted remove it with a mild detergent and wipe it with a dry cloth Chamber If the rotor chamber is not dry wipe moisture from the chamber with a dry cloth If th...

Page 15: ...tely Lid cannot be opened Press the Lid open button Lid cannot be closed Remove the dirt at the door latch and close the lid Check the lid latch is not damaged Unusual noise and vibration Check the device whether it is installed on the hard and flat place Reload the rotor symmetrically Reload the tubes symmetrically Tighten the clamping of the rotor with wrench by turning clockwise www ihanil com ...

Page 16: ... Low speed Deficiency not reaching 200RPM within 2 secs Check controller and motor Switch power off and re start E5 Door open Start Door open during operation Check door in switch E6 Under speed 2 under the set RPM during operation Switch power off and re start E7 System error Error of controlling system Overload of motor damage of IPM module Switch power off and re start contact Hanil A S center ...

Page 17: ...ntrifuge lid cannot be opened active the emergency release Switch off the power switch Wait for the rotor to stop completely Find the emergency release hole at the left side of the decvice Put a flat headscrewdriver into emergency release hole and turning clockwise www ihanil com 16 ...

Page 18: ...us MAX RCF MAX R mm MAX RPM Rotor ID 용량 Rotor A0 2M 32 A2 0M 18 A2 0M 24 32 x 0 2ml 18 x 1 5 2 0ml 24 x 1 5 2 0ml 2 3 1 12 500 15 000 15 000 9 695 17 910 21 055 55 5 71 2 83 7 Angle Rotors A2 0MP 24 Plastic lid available techsupport ihanil com 17 ...

Page 19: ...7 Rotors and accessories 7 2 Rotors and accessoriesfor Smart 15 Plus www ihanil com 18 ...

Page 20: ...x 30 Tube per Rotor 14 087 16 602 56 66 6 x 21 8 x 37 Black Black TR0 2 TR0 5 Tube Capacity ml 1 5 2 0 Dimensions Ø x L mm 11 x 38 11 x 38 Adapter Tube Type 18 Max RCF radius mm boring Ø x L mm 11 x 38 11 x 38 71 2 17 910 Color Material Cat NO Max height for tubefit 11 x 45 techsupport ihanil com 19 7 2 Rotors and accessoriesfor Smart 15 Plus ...

Page 21: ...8 x 30 Tube per Rotor 17 321 19 747 68 5 78 5 6 x 21 8 x 37 Black Black TR0 2 TR0 5 21 055 83 7 Dimensions Ø x L mm 11 x 38 Tube Capacity ml 1 5 2 0 11 x 38 11 x 38 Tube Type radius mm 24 Adapter Cat NO Color Material Max RCF boring Ø x L mm 11 x 38 56 mm Max height for tubefit www ihanil com 20 7 2 Rotors and accessoriesfor Smart 15 Plus ...

Page 22: ...MEMO ...

Page 23: ...Hanil Scientific Inc 16 Arayukro Gimpo 10136 Rep of KOREA T 82 2 3452 8965 info ihanil com www ihanil com ...

Reviews: