1. General Considerations

Follow precautions and all the safety requirements described on this user manual to prevent any

damages, failure of equipment and loss of lives.

1. The centrifuge should be installed on flat surface to maintain level.

     Abnormal vibration is caused when the device is installed on an inclined surface.
2. Check the voltage to be used, before connecting the centrifuge to the power source.

3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible    

    for damages of the device and accidents caused by using parts and accessories 

    not recommended.  
4. Do not exceed the maximum rated speed of the rotor or buckets or accessories in use.
5. Make sure to prepare necessary safety measures before using samples that are toxic or radioactive samples 

    or pathogenic or samples or infectious blood.

6. If the centrifuge is contaminated by toxic or radioactive samples or infectious blood samples, remove   

    contaminants completeyly and take needful actions such as ventilation or isolation of centrifuge.  
7. Substances that may generate volatile or explosive vapor can not be centrifuged.
8. The balancing work of samples shoud be done in advance before operation.
9. Before operation, rotor and chamber should be dry.
10. Do not attempt to slow or stop the spinning rotor by hand.
11. Only centrifuge with rotor and rotor lid firmly tightened.
12. Do not block vents.
13. Do not insert foreign objects into the holes of the device.
14. Repairs must be carried out by a technologist authorized by Hanil scientific Inc.
15. According to IEC61010-2-020 maintain a 30cm“clearance envelope”around the centrifuge while 
      the rotor is spinning.
16. Please contact the place of purchase or Hanil Scientific Inc. ([email protected]/[email protected])
      for product repairs.
17. Turn the power switch off after using the device.
18. Unplug the power plug before cleaning or left unused for a long period of time.

1.1 Safety

Summary of Contents for Smart 13

Page 1: ...Operation Manual Smart 13 Micro Centrifuge Date of purchase Serial No...

Page 2: ...opyright 2018 Hanil Scientific Inc All rights reserved Contact Us If you have any questions contactHanil Scientific Inc or place of purchase techsupport ihanil com info ihanil com Doc No OMM4PDEN1801...

Page 3: ...Package 6 2 3 Technical Specifications 7 3 Installation 8 3 1 Packing Inspection 8 3 2 Installation 8 3 3 Connect Power 9 3 4 Opeing Closing the lid 9 3 5 Loading and Removal of the Rotor 10 3 6 Loadi...

Page 4: ...the centrifuge is contaminated by toxic or radioactive samples or infectious blood samples remove contaminants completeyly and take needful actions such as ventilation or isolation of centrifuge 7 Sub...

Page 5: ...ontact your place of purchase or Hanil Scientific Inc Unusual noises or smell from the device Damage or wear of a power cord Breakdown of circuit breaker fuse or safety device If you spill liquid on t...

Page 6: ...ss the button to open lid Control Panel Keys and knobs for operating the centrifuge Power socket Socket is where the power cable connected Power switch Switch allows the device to be turned on or off...

Page 7: ...x speed 8 sec Tube Capacity 1 5 2 0ml x 12 Power supply AC110 220V 50 60Hz Dimension W x D x H mm 189 7 x 202 x 129 Weight kg 1 5 Cat No SM 13 RCF Table Rcf value was calculated based on the bottom of...

Page 8: ...be bent due to the weight of the rotor Good ventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidi...

Page 9: ...with a protective earth conductor and provided power cord Press the lid down until it clicks When lid is not closed completely the device does not operate O X 1 Connect the power cable to the power s...

Page 10: ...it completely out with the washer under the screw 2 Lift up the rotor from the motor shaft 3 Load the rotor into the motor shaft 4 Align the washer with the hole at the top of the rotor 5 Insert the...

Page 11: ...e same type of tube Tubes should be loaded symmetrically Do not exceed the maximum rated speed of the tube Same volume and density of sample should be put on tubes Check symmetric loading by balncing...

Page 12: ...lue Time setting knob Rotate clockwise or counterclockwise to set the run time To use pulse function set the time to 0 and press the start key Start Stop button Start or stop the centrifugation 6K Run...

Page 13: ...nterclockwise to set the run time 0 15 min 1 After setting the RPM and run time close the lid and press the start button If the lid is not completely closed it will not operate When the set time is ov...

Page 14: ...it turns smoothly 3 Use the stopwatch to check that the time setting is correct 4 If you find any damages do not use the device Contact Hanil Scientific Inc 5 Be careful not to get any foreign substa...

Page 15: ...rror sound Lid cannot be opened Press the Lid open button Lid cannot be closed Remove the dirt on the lid latch and close the lid again Check the lid latch is not damaged Unusual noise and vibration C...

Page 16: ...mm 44 04 54 19 59 06 Max RPM 13 500 Max RCF xg 8 973 11 042 12 032 Tube Tube capacity 0 2 ml 8 tube PCR strip Radius mm Inner line 31 13 44 06 Outer line 46 13 55 68 Max RPM 6 000 Max RCF xg Inner lin...

Page 17: ...MEMO...

Page 18: ......

Page 19: ......

Page 20: ...RECYCLABLE...

Reviews: