3. Installation
3.1 Packing Inspection
3.2 Installation
· Check packing conditions very carefully before unpacking.
· Contact Hanil Scientific Inc. immediately if damages found.
· Check the delivery for completeness.
Installation on hard and flat ground.
• Centrifuge should be installed on hard and flat place.
• If the centrifuge is installed in an inclined place, the shaft may be bent due to
the weight of the rotor.
Good ventailation.
• For air circulation and safety, maintain a 30cm“clearance envelope”around
the centrifuge while the rotor is spinning.
Constant temperature/humidity
• Centrifuge equipped with the sensitive electronic software which is fragile with
humidity and temperature.
• Must avoid direct ray or heater and be put in the ambience of controlled temp. and
air.
• Install the centrifuge in a place where corrosive gas is not generated.
• Sulfur dioxide gas and chlorine gas may cause corrosion.
Leveling
8 /20 www.ihanil.com
· You can get contact details on packing box or the bottom of the manual.
• The shaft should be put exactly vertical on the horizontaliy flat ground by
the leveling tool.
Avoid corrosive gas