background image

Operation Manual

 

Micro Centrifuge

Date of purchase

Serial No.

M 13

Summary of Contents for M 13

Page 1: ...Operation Manual Micro Centrifuge Date of purchase Serial No M 13...

Page 2: ...Only Contact Us Copyright 2020 Hanil Scientific Inc All rights reserved If you have any questions contact Hanil Scientific Inc or place of purchase techsupport ihanil com sales ihanil com UM M13 E Re...

Page 3: ...cifications 7 3 Installation 8 3 1 Packing Inspection 8 3 2 Installation 8 3 3 Connect Power 9 3 4 Opeing Closing the lid 9 3 5 Loading and Removal of the Rotor 10 3 6 Loading Tubes 11 4 Operation 12...

Page 4: ...d 6 If the centrifuge is contaminated by toxic or radioactive samples or infectious blood samples remove contaminants completely and take needful actions such as ventilation or isolation of centrifuge...

Page 5: ...on cord is needed make sure it its suitable for the device 4 Do not place anything on the power cable 5 Do not block vents 6 If you have the following emergencies shut off the power supply and unplug...

Page 6: ...on to open lid Control Panel Keys and knobs for operating the centrifuge Power socket Socket is where the power cable connected Power switch Switch allows the device to be turned on or off Absorbent r...

Page 7: ...alculated based on the bottom of the rotor hole RCF value can differ from the length of the tube inserted in the rotor AC110 220V 50 60Hz Weight without rotor kg 2 3 Technical Specifications 6 000RPM...

Page 8: ...y maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidity Centrifuge equipped with the sensitive electronic software which is fragile with hu...

Page 9: ...Press the lid down until it clicks When lid is not closed completely the device does not operate O X 1 Connect the power cable to the power socket on the back side of the device and plug the cord int...

Page 10: ...the washer under the screw 2 Lift up the rotor from the motor shaft 3 Load the rotor into the motor shaft 4 Align the washer with the hole at the top of the rotor 5 Insert the rotor lock screw into t...

Page 11: ...n matters 2 Tubes must be loaded symmetrically Always use the same type of tube Tubes should be loaded symmetrically Do not exceed the maximum rated speed of the tube Same volume and density of sample...

Page 12: ...20 www ihanil com M13 Rotate clockwise or counter clockwise to set the run time To use pulse function set the time to 0 and press the start key Rotate clockwise or counter clockwise to set the RPM val...

Page 13: ...eration When the lid opens during operation operation will stop 1 Rotate Time setting knob clockwise or counter clockwise to set the run time 0 15 min 1 Starting the centrifuge with setting the time k...

Page 14: ...moothly 3 Use the stopwatch to check the time setting for accuracy 4 If you find any damages do not use the device Contact Hanil Scientific Inc 5 Be careful not to get any foreign substances on the se...

Page 15: ...not be closed Unusual noise and vibration Check the device whether it is installed on the hard and flat place 6 Troubleshooting 6 1 General Errors 15 20 Remove the dirt on the lid latch and close the...

Page 16: ...M 13 500 Max RCF xg 8 965 11 003 12 032 Tube Tube capacity 0 2 ml 8 tube PCR strip Radius mm Inner line 31 13 44 06 Outer line 46 13 55 68 Max RPM 6 000 Max RCF xg Inner line 1 773 Outer line 2 241 Tu...

Page 17: ...MEMO...

Page 18: ......

Page 19: ......

Page 20: ...RECYCLABLE Doc No UM M13 E Rev 0...