Connect the device to voltage 
sources, which correspond to the 
electrical requirements on the label 
attached to the device.

Check the wall outlet is earth-
grounded, surge-protected and 
can supply enough power.

Connect the power cable to the 
distribution box, or install the plug 
and socket that are packed with 
the device.

Switch off the device after use.
Unplug the power plug before
cleaning or unused for a long
period of time.

Check the Line Voltage

Grounded Outlet

Power Connection

Switch Off the Device

Quick Guide

Component wR

Cat. No. : CP-wR

WS750-6 

6 x 750 mL / Max. 4.5k

30 cm

30 cm

30 cm

Allow a safety clearance of 
30 cm around the centrifuge 
during operation and 
storage.

Care of Swing-Out 
Rotor

Remove Dust

Empty the Condensation 
Water

Allow a Safety Clearance 
of 30 cm

Emergency Lid Open

If the centrifuge lid cannot be opened,
insert the provided T-wrench into the 
emergency release hole on the right side 
of the device and press it. (Before activat-
ing emergency open, switch off the
device and wait for the rotor to stop.)

Spread the grease lightly 
on the bucket where it 
contacts the rotor and the 
pivot.

Remove any dust from the 
ventilation slits on the right 
side of the device.

Empty the condensation water 
regularly by removing the 
drain cap on the left side of the 
device.

3.0KVA
220~240V 50/60Hz 

Maintenance

Electrical Safety

Rotors

Cautions for Use

Before Operation

After Operation

Check the rotor, buckets and tubes are free 
of scratch and crack before operation.

Check the Accessories for Abnormalities

After relocating the device or loading the rotor, 
balance them by using a leveler. If necessary, 
level the device by adjusting the ‘Leveling foot’ 
on the bottom of the device.

Balancing the Centrifuge and Rotor 

Load tubes symmetrically into the rotor  
with identical amount.

Load Tubes Symmetrically

Place the lid on the wind shield rotor before 
centrifugation. 

Check the Lid

Wipe up the moisture in the chamber with a 
dry cloth after use. Keep the lid open when 
not in use.

Clean the chamber and rotor by wiping with 
a cloth soaked in mild detergent and dry 
completely. 

Keep the Chamber Dry

Cleaning

Rev.1

Proudly made by Jiyoung Shim, Hyewon Jeong 

Order & Inquiry l +82-2-3472-0727  l [email protected] l www.ihanil.com
Service  l +82-2-3452-8966 l [email protected]