background image

Operation and Maintenance

1. Place the Porta-Vac on a sturdy work surface.
2. Attach the power cord to the receptacle on the  

  unit and plug the other end into a power outlet  

  following the safety guidelines in this manual. 
3. The power switch (shown at right) is located  

  at the top corner of the fan door turns on the fan  

  motor and LED light. The DC cable must be  

  plugged in to use the LED light. 
4. Clean and wipe the polycarbonate glass shield  

  with a soft cloth or paper towel. A non- 

  ammoniated cleaner (such  as soap and water)  

  will remove greasy polishing compound dust. 
5. The Porta-Vac has been fitted with a fine pleated  

 HEPA 

WASHABLE

 filter. Periodically check the  

  filter condition. If it becomes clogged with debris  

  you can remove it by opening the rear door of  

  the fan housing. You can either tap it on its side  

  in a trash can to gently remove large particles  

  and return it to its original position in the  

  chamber or, if it is heavily clogged, replace the  

  filter with a new one. Please make sure that the  

  filter is inserted all the way into the  

  compartment. Filters can be saved and sent to  

  a refiner to recover precious metals. The pleated  

  filter (Part No. 550-F) should be replaced every  

  4-6 months   under normal use and more often  

  under heavy use. Contact your distributor or  

  Handler Manufacturing LLC  to request a  

  550-F pleated  filter.
6. To clean and remove the dust which  

  accumulates inside the Porta-Vac, lift  the shield  

  as shown on the right. Remove the tray, and  

  place tray over the trash and empty contents.
7. With use, the shield of the unit may become  

  cloudy looking or scratched by debris. When  

  this occurs Contact your distributor or  Handler  

  Manufacturing LLC to request a replacement of  

  shield P550-04P

AC receptacle  

with power cord 

removed

Pry open  

fuse drawer

Active 

Fuse 

Spare 

Fuse

8. The Porta-Vac has a Bussman GMC-1A fuse  that powers the fan and the light. The fuse and a spare fuse 

are stored in the AC power cord receptacle near the rear of the chamber. 

To check or change the fuse, follow these directions: 

Remove the power cord from the AC receptacle to 

expose the fuse holder drawer. Use the blade of 

a small screwdriver in the slot of the fuse holder 

and gently pry outward. The active fuse is in the 

back of the drawer and a spare fuse is in the front 

compartment. If the glass case of the active fuse is 

darkened, the fuse is blown. From the underside of 

the drawer, push out the blown fuse using a small 

screwdriver.
Do not use a higher ampere fuse.

DC Cable

Power Switch

Washable  
HEPA filter  
Part # 550-F

Tray  
Part # P31-07A

Shield 
Part # P550-04P

Power 

Receptacle and 

Cordset

Reviews: