General Inspection and Maintenance
C) Annually - To be completed by a lift technician
Annually,
the
C450/ C625
system
must be inspected. It is recommended that this service history is
documented; see pages 34-36 for Service Record History templates.
□
Complete the inspection as noted in the “Monthly” section above.
□
Complete the preventative maintenance procedure as outlined in technical manuals for the
C450/ C625
system
.
D) Semi-Annually - To be completed by User
In high frequency transfer areas (more than 1500 lifts per year) or situations where heavier than normal clients
regularly are lifted, PM should also be completed every 6 months. It is recommended that this service history
is documented; see pages 34-36 for Service Record History templates.
Should any of the these items fail the inspection do not use the lift. Contact your local authorized
dealer for service.
Check for the following:
□
Check emergency stopping and emergency lowering functions using the RED emergency cord
□
Inspect carry bar for damage; verify insert and hooks. Ensure no cuts, dents, or sharp edges exist
that may damage the strap of the unit or sling
□
Lower the lifting tape all the way out of the unit; verify NO signs of fraying or breaking along its
entire
length.
□
Remove the plug from the carry bar and inspect the integrity of the strap around the pin.
□
The stitching on the lift lifting tape where it connects to the carry bar shows NO signs of fraying,
or
breaking.
□
Verify all functions on the hand control work correctly (UP, DOWN, TRAVERSING, RETURN
TO CHARGE etc.)
□
The lift has no unusual sounds when the carry bar is moved UP/DOWN or the lift is moved LEFT/
RIGHT.
□
The airline tube that connects the hand control to the lift is not kinked, twisted, knotted, leaking,
cut or damaged.
□
Check that the grommet connectors are tight
□
Hold UP until the carry bar is at the top. Motor should stop automatically.
□
Using the hand control, let the strap out all the way to the ground. Motor should stop before strap
winds backwards.
□
While using either UP or DOWN, angle the strap more than 15 degrees. Motor should stop
automatically.
□
Engage unit with charger and check that unit is charging properly. Light on the lift should turn
amber.
□
Check that the lift’s LED light turns green when on, dark when off, and amber when charging.
□
Check for cracks or wear in case
Note: if the audible/visual preventative maintenance alert sounds, complete the PM outlined in Section D prior to reset-
ting the lift counter. Instructions to reset the lift counter are detailed on page 21.
This section to be only completed by a qualified service technician as authorized by Prism
Medical.