V. MAINTENANCE
MONTHLY - Depending on use
The bill path, also known as the platen assembly, should be opened and cleaned regularly. To do this:
1.
Unplug the changer.
2.
Disconnect the 9-pin plug.
3.
Remove the Validator from the holder.
4.
Open the bill path (Refer to Appendix B for instructions).
5.
Clean rollers, heads, belts and sensors with cotton swabs and rubbing alcohol.
6.
Close the platen assembly.
YEARLY
The STA Validator should be serviced annually to maintain maximum performance. THIS WORK SHOULD
ONLY BE DONE BY A TRAINED TECHNICIAN.
STA Validator Calibration
1)
Take a clean piece of white copier paper and cut it to the size of a bill (test bill).
2)
Set the switches on the side of the validator so that only #8, #7 and #1 are in the
ON position.
3)
Insert the paper. It should step in slowly (takes about 8-10 seconds).
4)
The validator will reject the paper. When you remove the paper from the validator,
the display should flash “11”. This means the process was completed.
5)
Turn all the switches back to the OFF position.
6)
IMPORTANT! Power down the validator for about 15 seconds. Then power it
back up and try a bill.
7)
Calibration complete.