background image

Subject to change without notice

22

or CHP)

 displayed in the readout can be overwritten

and changed to the opposite mode. This is carried out
by simultaneously pressing and holding the

 CHI (22) 

and

the 

DUAL (23) 

pushbutton.

STORAGE MODE ONLY

The time deflection ranges are different in storage mode
and therefore must be considered (

please note item

(7)

). In 

DUAL

 mode a selection between 

alternate

 or

chopped

 signal capture is 

neither required nor

possible

. Consequently the readout does not display

such information 

(ALT, CHP)

.

In 

XY

 storage mode the 

TIME/DIV.

 knob is still active.

It is used to determine the sampling rate, which is then
displayed by the readout.

(30) A/ALT-

B - 

Pushbutton with multiple function.

The instrument contains two time bases designated 

A

and 

B

. With the aid of the 

B

 time base, signal parts

displayed by the 

A

 time base can be expanded in X

direction. The expansion ratio depends on the time
deflection coefficient ratio of both time bases (i.e. 

“A:

100µs”

,

  “B:1µs” = 100

). With higher expansion ratios

the 

B

 time base trace intensity reduces if analog mode

is active.

SINGLE (SGL)

 event signal capture mode is enabled in

A

 time base mode only.

A/ALT 

function

:

Each time the 

A/ALT- B

 pushbutton is briefly pressed,

the time base mode changes in the sequence 

A

  –

alternate 

A

.

A

In 

A

 time base mode the

 TIME/DIV. (29) 

control knob

is operative only for this time base. The readout then
only displays the 

A

 time coefficient. The time base

settings for this condition are stored if the time base
mode is changed.

ALT

If 

alternate

 

(ALT)

 time base mode is selected, the

TIME/DIV (29) 

knob only controls the 

B

 time base.

The alternate time base mode is a sub function of the 

B

time base mode where both time base traces are
displayed. Consequently the readout displays both time
deflection coefficients (e.g.

  “A:100µs B:1µs”

). Unlike

the former 

A

 time base mode, an intensified sector is

also visible on the 

A

 trace. This sector indicates the signal

part which is displayed by the 

B

 time base. The

intensified segment can be shifted horizontally by the

DEL. POS. (28)

 control knob continuously (if the B time

base is operated in free run conditions). The difference
between the start of the 

A

 time base trace and the

beginning of the intensified sector shows the delay time.
This information is also displayed in the readout as an
approximate value (e.g. 

“Dt:2.5ms”

) related to the

calibrated 

A

 time coefficient (uncalibrated i.e. 

“Dt>

2.5ms”

). The width of the segment decreases when

the B time coefficient is set to a lower value (higher
time deflection speed).

Alternate time base mode causes the display to alter
between 

A

 and 

B

 time base after each sweep (analog

mode). In alternate

 DUAL 

mode the sequence is channel

I with

 A 

time base,

 channel I 

with

 B 

time base,

 channel

II 

with

 A 

time base and

 channel II 

with

 B 

time base.

For better reading, the vertical position of the 

B

 trace

position can be shifted (

please note “TRS” (13)

).

function

:

Pressing and holding the pushbutton switches over from

A

 or 

alternate

 time base mode to 

B

 time base mode

and vice versa.

In 

B

 time base mode the display of the 

A

 trace(s), the

intensified sector(s) and the 

A

 time coefficient display

in the readout are disabled. As the trace separation 

(TRS)

is no longer required under these circumstances, this
function is switched off too. Consequently only the 

B

time coefficient is displayed by the readout.

Briefly pressing the pushbutton switches 

B

 time base

mode off and 

A

 time base mode on. Pressing and holding

the pushbutton switches over from 

B

 to 

alternate 

time

base mode.

STORAGE MODE ONLY

In the combination of storage mode and 

alternate 

or 

B

time base operation, the pre- and post trigger are
automatically switched off. The

 PTR (8) pushbutton

 is

deactivated and instead of a pre- or post trigger value
the readout displays the delay time.

ROLL 

mode is not available in 

alternate

 or 

B

 time base

mode.

(31) DEL. TRIG.

VAR. - 

Pushbutton with double function.

DEL. TRIG. 

function

:

ANALOG MODE ONLY

Each time the pushbutton is briefly pressed, the instrument
switches between free run (untriggered) and triggered 

B

time base, if 

alternate

 or 

B

 time base mode is present.

DIGITAL MODE ONLY

In contrast to analog operation the 

B

 time base can not

be triggered in 

alternate

 time base mode.

ANALOG and DIGITAL MODE

The actual setting is displayed in the readout. Instead of
the approximate delay time (

“Dt:...”

) in free run mode,

the readout displays 

“DTr: slope, DC (trigger coup-

ling)”

 in triggered delay mode. In this mode, the former

A

 time base trigger settings regarding the trigger mode

(automatic or normal), -coupling, -slope and -level
settings are stored but still active.

With the activated delay trigger the instrument is
automatically set to normal 

(NM)

 trigger mode and 

DC

Controls and Readout

Summary of Contents for HM1507-3

Page 1: ...ls 32 Line Mains triggering 32 Alternate triggering 32 External triggering 33 HOLD OFF time adjustment 33 B time base 2nd time base 33 Triggering after Delay 33 AUTO SET 34 Component Tester analog mod...

Page 2: ...operation the following hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer...

Page 3: ...Variable 2 5 1 up to 1 25s div X MAG x10 5 max 5ns div Holdoff time variable to approx 10 1 Bandwidth X amplifier 0 3MHz 3dB X Y phase shift 3 below 220kHz Digital Time Base Accuracy 3 1 2 5 sequence...

Page 4: ...in a socket outlet provided with a protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor The mains line plug must be insert...

Page 5: ...on ments The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be use...

Page 6: ...wnwards at each change Pure direct voltages can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbo...

Page 7: ...In this case an appropriate high voltage capacitor approx 22 68nF must be connected in series with the input tip of the probe With Y POS control input coupling to GD it is possible to use a horizontal...

Page 8: ...its full length at not too great steepness and that the horizontal distance at 10 and 90 of the amplitude is measured If the edge shows rounding or over shooting the 100 should not be related to the p...

Page 9: ...as short and thick as possible When the attenuator probe is connected to a BNC socket a BNC adapter should be used In this way ground and matching problems are eliminated Hum or interference appearing...

Page 10: ...and no other LED on the front panel is lit except A All INTENS settings are stored after the instrument is switched off The AUTO SET function switches the readout on and selects A time base mode A LED...

Page 11: ...cally again it will be required to stop rotating the knob for at least about 2 seconds Attention The dynamic range limits of the A D converter may become visible if a Y position shift is performed af...

Page 12: ...STOR MODE Pushbuttons with associated LEDs These functions are not available in analog mode If digital SINGLE SGL mode has not been chosen one of the associated LEDs is lit The signal capture and disp...

Page 13: ...g mode on or off SINGLE mode is indicated by the associated SGL LED SINGLE mode is available in digital as well as in analog mode and remains unchanged when switching over from analog to digital mode...

Page 14: ...tton until a beep confirms reference memory acquisition Before starting the procedure the current signal can but must not be protected by the HOLD function As the reference signal is then equal to the...

Page 15: ...the DC content of a signal by comparing the actual Y position with the previously determined 0 Volt Y position Y POS I Symbol Provided that the readout is displayed and DC REFE RENCE ON is selected i...

Page 16: ...control knob causes a different trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal voltage crosses the trigger point In most Yt modes the trigger po...

Page 17: ...o GD 34 38 Whether alternated or chopped channel switching is present in analog mode depends on the actual time base setting and is displayed in the readout ALT analog mode only displayed in the reado...

Page 18: ...e A B time base mode B time base mode and line mains trigger coupling STORAGE MODE ONLY In ROLL mode the signal capture is untriggered Con sequently alternate triggering is not available 25 VOLTS DIV...

Page 19: ...start of the intensified sector In the free run condition delay trigger not active of the B time base an approximate value is displayed in the readout Dt This is an aid to find the position of the int...

Page 20: ...information is also displayed in the readout as an approximate value e g Dt 2 5ms related to the calibrated A time coefficient uncalibrated i e Dt 2 5ms The width of the segment decreases when the B t...

Page 21: ...on coefficient setting is not possible in storage mode Underneath the front panel sector described before the BNC sockets and four pushbuttons are located 32 INPUT CH I BNC socket This BNC socket is t...

Page 22: ...he CRT there are the controls for the readout the component tester and the square wave calibrator with their outputs Please note The following description of the cursor related controls assumes that t...

Page 23: ...nt must be activated Briefly pressing the CHI II 1 t ON OFF pushbutton switches from f to t and the following measure ment modes until the readout displays f again In all conditions vertical cursor li...

Page 24: ...ayed in the readout must be multiplied by 100 if e g a x100 100 1 probe is used In case of x10 10 1 probes the probe factor can be automatically included see AC DC 33 and AC DC 37 t In time or frequen...

Page 25: ...connect the test lead s to the test point s Then switch on the instrument and after wards the device under test The oscilloscope is switched on by depressing the red POWER pushbutton After a few seco...

Page 26: ...tton is depressed Connect the probe to CH I input Depress the CAL pushbutton for 1MHz Operate the oscilloscope as described under 1kHz but select for 0 2 s div time deflection coefficient setting Inse...

Page 27: ...are the same for both channels because the X x10 magnifier is inactive in XY mode The bandwidth of the X amplifier is lower than the Y amplifier and the phase angle which increases with higher freque...

Page 28: ...l cycles of the modulation frequency are visible Strictly speaking triggering should be external with modulation frequency from the AF generator or a demo dulator However internal triggering is freque...

Page 29: ...ger coupling setting In the absence of an adequate trigger signal or when the trigger controls particularly the trigger LEVEL control are misadjusted no trace is visible i e the screen completely blan...

Page 30: ...ernal trigger voltage itself is displayed first with internal triggering In most cases the composite video signal has a high DC content With constant video information e g test pattern or color bar ge...

Page 31: ...iting the trigger pulse This causes an untriggered display and a flashing trigger LED HOLD OFF time adjustment For instrument specific information please note DEL POS HO 28 in section Controls and Rea...

Page 32: ...ically for a display of approx 2 signal periods The time base setting occurs randomly if complex signals consisting several frequencies e g video signals are present AUTO SET sets the instrument autom...

Page 33: ...mination of actual values will come with experience or by direct comparison with a component of a known value Testing Capacitors and Inductors Capacitors and inductors cause a phase difference between...

Page 34: ...ared with single components In case of doubt one component terminal may be unsoldered This terminal should then not be connected to the ground socket avoiding hum distortion of the test pattern Anothe...

Page 35: ...d Controls Vertical resolution The dot density in each operation mode is 8 bits 28 256 dots displayed over a height of roughly 10 divisions The instrument is adjusted for 25 dots per division This eas...

Page 36: ...k must on no account be visible Visible trace fault without input signal bright dot on the left side or decreasing brightness from left to right or shortening of the baseline Cause Incorrect Unblankin...

Page 37: ...mitted This is achieved by adjusting the trimmer located in the 2 1 attenuator The shape of the square wave should then be the same in each input attenuator position Operating Modes CH I II DUAL ADD C...

Page 38: ...Then both readout information regarding the time deflection coefficient and the time difference show exactly the same value After switching over from time to frequency measurement the readout display...

Page 39: ...ng on of the oscilloscope and the first command SPACE CR 20hex 0Dhex sent from the PC the baud rate is recognized and set automatically between 110 baud and 19200 baud The oscilloscope is then switche...

Reviews: