background image

 

 

4

 

 

The following describes the operational features of the L-708 Laser, shown 

in Figure 1.  

 

 

The power switch 

has a lighted LED to indicate that power is ON.  

 

The  battery  pack connector 

accepts a slip-fit probe with a flexible 

cord.  

 

The bubble level vial 

on  the laser mounting flange is  used for 

rotational accuracy. When the bubble  in the level vial is  centered 

horizontally, all adjustments controlling laser beam angle will shift the 

laser beam vertically or horizontally with reference to the bore/target 

axis. If the bubble is 

not 

centered, any adjustment to one laser axis will 

change the laser beam position in both axes. The levels also provide 

fixture mounting repeatability.  

 

Precision Angular Adjustment Knobs 

provide angular pointing control for the laser beam so it can 

be tilted with a resolution of .0007 in. in 100 ft. (0.018mm in 30.5 m). The L-708LD laser has an 

angular adjustment resolution of .0014  in.  in 200 feet 

(0.036 mm in 61 m).  

 

The  Center of Adjustment Range Indicator for 

Coarse Adjustments

 is a small window containing 

a white dot. The white dot follows the coarse 

angular adjustment of both axes.   If the white dot 

displays  in the center of the window, both the 

Horizontal and Vertical angular adjustments  are 

centered. 

 

The Fine Adjust Tool Bar 

swivels to indicate the 

middle of the adjustment range of the fine 

adjustment knobs.  Rotate the bar (located under the 

Vertical and Horizontal Fine Adjust knobs)  and 

tighten  the knobs until they nearly touch the tool 

bar. Rotate the bar away from the knobs and the 

knobs are now in the middle of the adjustment 

range. 

 

 

Figure 3 – 

Fine Adjust Tool Indicator in the OUT position 

 

Starting an Alignment with the L-708 

To start the alignment, it is important to make sure the L-708’s angular adjustments are in the center of 

their adjustment range.  There are two indicators to aid in setting the adjustments in the center of their 

range: 

 

1.

 

Coarse Angular Adjustment 

 

Indicator

 – a white dot in a round window 

that indicates how far off center the coarse 

angular adjustments are.  To start an 

alignment, ensure the white dot is centered 

in the window using the 

coarse

 angular 

adjustments.   

 

 

 

Figure 2 – 

Top of L-708 Laser  

showing bubble level vial 

Summary of Contents for L-708

Page 1: ......

Page 2: ...ticles of equipment manufactured by other companies and limited by a warranty extending for less than one year In such an event the more limited warranty applies to said instrument or article of equip...

Page 3: ...S and A 514GL 11 Setting the Legs for the A 514GS Using Small and Medium Bore Adapters 12 Setting the Legs for the A 514GL Using Large Bore Adapters 13 The Model R 1307 Readouts 14 Model R 1307C 14 Mo...

Page 4: ...ing error of up to 002 in 0 05 mm can occur In addition the laser switch setting CONT or Fixed vs PULSE must also agree with the R 1307 Readout setting For CONTinuous Fixed Mode set the readout to F10...

Page 5: ...to the desired bore radius using the Leg Setting Gage The L 708 Laser and A 512 Target are inserted into their Bore Adapters The entire laser and target assemblies are inserted into the two reference...

Page 6: ...der good conditions should hold 010 0 25 mm over the full distance The L 708LD requires the use of the T 1218 2 Axis Target T 1220 or T 1240 targets 20x20 mm PSDs since the laser beam diameter is larg...

Page 7: ...Adjustments is a small window containing a white dot The white dot follows the coarse angular adjustment of both axes If the white dot displays in the center of the window both the Horizontal and Vert...

Page 8: ...t establishing a reference bore centerline Pulse Continuous Modes for the L 708 Laser The L 708 Laser is equipped with a PULSE CONTinuous switch which manually switches the laser beam between Pulsed a...

Page 9: ...and remove the cover 3 Replace the two batteries Remove the old batteries and replace them with new 9 volt cells being careful to orient them with the negative terminal out up 4 Re attach the cover R...

Page 10: ...he counter weight to the poles 4 An extra counter weight and pole are included with the A 514C Insert the pole in the main counterweight and use the thumbscrew to tighten 5 Slide the extra counterweig...

Page 11: ...t and adapter in self centering mode are inserted into the reference bore A measurement is taken with the target in both the NORMal and INverted positions see Appendix A The NORMIN Method beginning on...

Page 12: ...for bores from 6 5 in 165 mm to 17 5 in 445 mm A 514C for bores from 17 0 in 432 mm to 40 0 in 1 m Note For bores over 1 m half bores or where bore surfaces are worn or rough the T 218 Two Axis Univer...

Page 13: ...e Adapters The A 514A B and C Bore Adapters are all assembled the same way Each adapter consists of two legs that cross each other forming an X shape Using a bolt and Allen wrench provided screw the b...

Page 14: ...S is used with the small A 514A or medium A 514B bore adapters The A 514GL can be used for all three adapters and must be purchased if using the large bore adapter Figure 6 The A 514GS left and A 514G...

Page 15: ...e the stop pin and turn the bore adapter one more time replacing the stop pin Repeat the process for the next leg 9 To set the remaining two legs the bore adapter must be removed from the hub and flip...

Page 16: ...e other end in the slot 3 Set the caliper to 7 in and press one end of the caliper against the pin of the extension 4 Slide the leg setting plate up against the other end of the caliper 5 After the le...

Page 17: ...07C Supports cabled local targets only Supports both pulsed beam and continuous laser modes Functional replacement for the R 307 Analog Readout Model R 1307 900 2 4 and R 1307 2 4ZB Supports both wire...

Page 18: ...r beam to target position information and plot results that can be saved and printed The R 358 computer interface provides exceptionally high accuracy 00004 in resolution for downloading target data t...

Page 19: ...16 Model R 1307 Readout Control Panel...

Page 20: ...17 Model R 1307C Readout Control Panel...

Page 21: ...a contraction of NORMal INverted which briefly describes the method In Step 3 Laser Setup on screen instructions guide the user through setting up the laser and making it parallel to reference points...

Page 22: ...re provided with counterweights to hold the laser in place so the unit doesn t tip 3 Insert the laser and target assemblies into the two reference bores Attach the weighted insertion rod to the target...

Page 23: ...il the upper display shows Use the UP and DOWN arrow keys to select 5 Select the PSD descriptor applicable to your target Press the MENU button until the upper display shows Use the UP and DOWN arrow...

Page 24: ...E READOUT AS FOLLOWS R 1307C SET FUNCT TO LOCAL ID AND CH NOT USED R 1307 SET FUNCT TO LOCAL SET ID OF UPPER AND LOWER DISPLAY TO THE SAME VALUE DEFAULT 1 SET CH TO A VALUE NOT USED BY ANOTHER READOUT...

Page 25: ...ontal X axis position of the local target 4 Set the System ID Radio Channel Press the MENU button until the upper display shows with the current System ID blinking Use the UP and DOWN arrow keys to se...

Page 26: ...ed in bore alignments the True Bore Center the Target Center and the Laser Reference Centerline If mounting fixtures were perfect the Target Center would be located at the True Bore Center and if perf...

Page 27: ...on the readout are also rotated Therefore although the Target Center appears to be to the right of and lower than the Laser Beam Center in Figure 18 the vertical readings are positive and the horizont...

Page 28: ...sured point is the Measured Diameter TCE MDT The formula is MDT Normal Reading Inverted Reading 2 3 Calculate the difference in diameter between each measured point and the reference location The diff...

Page 29: ...are provided as well as a sample plot of readout data see Figure 20 on Page 27 LOCATION NORMAL INVERTED OFFSET DIAMETER Reference Location Measuring Point 1 Measuring Point 2 Measuring Point 3 Measur...

Page 30: ...27 Figure 20 Sample plot of readout data...

Page 31: ...01 in ft 0008 mm m Beam Stability 0001 in hr F 0 004 mm hr C 0001 in ft hr F 0 01 mm M hr C Laser Type Diode Laser 0 9 mW Cw BRH Class II Laser Wavelength 670 nanometers Materials Aluminum Flange All...

Page 32: ...29...

Page 33: ...the type intended for nail polish removal NOTE When cleaning precision optics even with the best quality optical cleaning tissue use gentle pressure to avoid scratching the surface or damaging the opt...