C11440-42U30/C11440-42U31 Instruction manual_Ver.1.2
20
■
Calculation formula (Rapid Rolling Mode)
Vn = Number of vertical line
(The center of the set area is the middle of the sensor.)
Exp1 = 1 ms to 10 s
Exp2 = 1.05 ms to 10 s
1H
= 10 × 10
-6
int () = The decimal point is rounded down.
•
The Exp1 and Exp2 values must be input to the calculation formula below in units of
seconds.
Binning
Horizontal
width
Vertical
width
Calculation formula
Free running mode
1×1
1024
1536
2048
---
1/(int(Vn/2048/30/1H)×1H)
1×1
512
---
1/(Vn/2×1H)
2×2
4×4
---
---
External trigger mode
(Edge trigger)
1×1
1024
1536
2048
112
≤
Vn
≤
2048
1/(int(Vn/2048/30/1H)×1H)
1×1
8
≤
Vn
≤
104
1/(Vn/2×1H+1H×10+Exp1)
1×1
512
---
2×2
4×4
---
---
External trigger mode
(Level trigger)
1×1
1024
1536
2048
112
≤
Vn
≤
2048
1/(int(Vn/2048/30/1H)×1H)
1×1
1024
1536
2048
8
≤
Vn
≤
104
1/(Vn/2×1H+1H×4+Exp2)
1×1
512
---
2×2
4×4
---
---
External trigger mode
(Synchronous readout)
1×1
1024
1536
2048
32
≤
Vn
≤
2048
1/(int(Vn/2048/30/1H)×1H)
1×1
64
1024
1536
2048
8
≤
Vn
≤
24
1/(Vn/2×1H+1H×18)
1×1
512
---
2×2
4×4
---
---