WI-25023
22
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E32]
▲
Stimulus Hardware Current Limit
Turn AC power OFF-ON to attempt
recovery. Contact technical support
if problem persists.
1. Carefully reposition and stabilize
cannula or introducer
2. Check probes, cables and
grounding pad connection
3. If problem persists, remove probe
from cannula/introducer, wipe off tip
and replace
4. Initiate new procedure and check
for proper function
5. Swap out probe, cable and
grounding pad as appropriate
6. Turn unit “OFF” then “ON” to
attempt recovery
7. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E33]
▲
Stimulus Hardware Voltage Limit
Turn AC power OFF-ON to attempt
recovery. Contact technical support
if problem persists.
Table of Error, Fault & Warning Codes
Code
Message
Recommended Action
(Complete List)
FAULT [E34]
▲
Stimulus Hardware Pulse
Duration Limit
Turn AC power OFF-ON to attempt
recovery. Contact technical support
if problem persists.
1. Turn unit “OFF” then “ON” to
attempt recovery
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
FAULT [E35]
▲
Software Error Detected
Turn AC power OFF-ON to attempt
recovery. Contact technical support
if problem persists.
1. Turn unit “OFF” then “ON” to
attempt recovery
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
FAULT [E36]
▲
Software Error Detected
Turn AC power OFF-ON to attempt
recovery. Contact technical support
if problem persists.
1. Turn unit “OFF” then “ON” to
attempt recovery
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3