WI-25023
29
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E106]
▲
Cooled Temperature Out-Of-
Range
Outside 18-34°C expected range.
Probe(s), cable(s) or pump unit may
be defective.
1. Check probe and cable
connections
2. Initiate new procedure and check
for proper function
3. Carefully reposition and stabilize
cannula or introducer
4. Swap out probe, cable and
grounding pad individually
5. If problem persists, remove probe
from cannula/introducer, wipe off tip
and replace
6. Try new probe(s) and cable(s)
7. Turn unit “OFF” then “ON” to
attempt recovery
8. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E107]
▲
Probe B Connected But Disabled
In Advanced Settings
1. Disconnect TDP B/Secondary
Thermocouple
2. If desired, enable TDP B/Secondary
Thermocouple in ADVANCED
SETTINGS
ERROR [E108]
▲
Probe A Not Connected
Check probe and cable
connections. Probe or cable(s) may
be defective. (TDP A)
1. Check probe and cable
connections
2. If desired, enable appropriate
Probe in ADVANCED SETTINGS
3. Try new probe(s) and cable(s)
4. Turn unit “OFF” then “ON” to
attempt recovery
5. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E109]
▲
Probe B Not Connected
Check probe and cable
connections. If desired, enable 1
PROBE in ADVANCED SETTINGS.
(TDP B)
ERROR [E110]
▲
Probe A and Probe B Not
Connected
Check probe and cable
connections. Probe(s) or cable(s)
may be defective.
(TDP A AND TDP B)
ERROR [E111]
▲
Probe A Connected to Wrong
Side of Y-Cable
Disconnect probe and attach to
other side of Y-cable.
1. Disconnect probe and attach to
other side of Y-cable
2. Check probe and cable
connections
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3