Appendix A:  Installation of Peristaltic Pump 

NOTE: Perma Pure peristaltic pumps are rated for a maximum pressure of 20 psi. 

If supplied loading key is not available, use a flat head screwdriver. 

 

Single pump head loading 

 
a)

 

Separate the end bells (the pump head halves). Hold the end bell containing the 
rotor as shown with the tubing retainer grooves facing down. 

 
b)

 

Place tubing in the right groove and against the first two rollers. Hold tubing 
with thumb. Near groove, insert smaller prong of loading key between the top of 
the rotor and tubing. Push key in as far as possible. 

 
c)

 

Push down and turn key counterclockwise completely around the rotor. The key 
will push the tubing uniformly into the end bell assembly. Hold the second end of 
the tubing. Remove key. 

 
d)

 

Position the other end bell on top and press the end bells together. Be careful 
not to pinch the tubing. If end bells do not snap tightly together, reload tubing. 
If necessary, turn key in slot on rotor shaft to adjust tubing (as in step e). 

 
e)

 

With key in slot on rotor shaft, turn key to align tang on rotor shaft with slot in 
motor drive shaft. Point tubing retainer grooves up. Shift the pump head slightly 
till it snaps on the alignment pins (if present). Secure with four provided screws. 
Tighten with fingers only. 

 

 

 

 

Multi-channel mounting 

 
a)

 

Load the pump heads with tubing. 

 
b)

 

Install the four correct length mounting screws in drive. 

c)

 

Slide the first pump head into the mounting screws. 

 
d)

 

Place key in slot on mounting shaft. Twist to align tang on rotor shaft with slot in motor drive shaft. Shift the pump 
housing around till it drops over the alignment pins (if present). 
 

e)

 

Repeat for each additional pump head, aligning pump head tang with slot on previously mounted pump head. 

 
f)

 

Slide the flow flat washers onto screws and secure with the four wingnuts. Tighten with fingers only. 

 

Replacement parts 

2PBM-001     Peristaltic pump, motor only, 115 VAC 60 Hz 

2PBM-002     Peristaltic pump, motor only, 230 VAC 50 Hz 
2PBM-003     Peristaltic pump, head only, standard 

2PTB-002PK  Peristaltic pump, tubing, Norprene, size 17 (10 feet) 
2PBM-006     Dual head hardware kit  

Summary of Contents for PERMA PURE Baldwin 4S-9PA

Page 1: ...A PURE LLC 732 244 0010 800 337 3762 fax 732 244 8140 www permapure com info permapure com PERMA PURE LLC 8 Executive Drive Toms River NJ 08755 Baldwin Series Sample Conditioning Systems Technical Man...

Page 2: ...nnections Error Bookmark not defined LED Summary 5 Start up 5 Set Points Error Bookmark not defined Shutdown Error Bookmark not defined Replacement Parts 8 Appendix A Installation of Peristaltic Pump...

Page 3: ...0410 Sample gas flow range Water content Maximum cooling rate Maximum inlet sample temp Maximum inlet pressure Maximum pressure drop Ambient temperature range Outlet sample gas dew point Voltage Power...

Page 4: ...PERMA PURE LLC Sample Conditioning Systems User s Manual Doc 580 Revision 000 Page 4 of 16 o Copy M Class Safety info...

Page 5: ...cators please verify that all of the connections to the Sample Conditioning System are correct Once this re check has been completed please energize Sample Conditioning System The unit should be ready...

Page 6: ...indicates ambient temperature Thermocouple placement in heat exchanger block Ensure proper placement Replace thermocouple or control board Impinger does not reach set temperature but is below ready te...

Page 7: ...use excess condensate and could lead to water slip Sample pump does not start Ready LED is on Slip LED is Off alarm relay water slip option installed Electrical connections to sample pump Verify that...

Page 8: ...3 Thermocouple Temperature Control Type K 36 3KFA 001 Complete filter assembly sample in line 2 micron 3FHG 001 Filter bowl glass 3FEC 002 Filter element ceramic 2 micron 3KPB 003 Peristaltic pump dua...

Page 9: ...not snap tightly together reload tubing If necessary turn key in slot on rotor shaft to adjust tubing as in step e e With key in slot on rotor shaft turn key to align tang on rotor shaft with slot in...

Page 10: ...PERMA PURE LLC Sample Conditioning Systems User s Manual Doc 580 Revision 000 Page 10 of 16...

Page 11: ...PURE LLC SE MAN 003 Sample Conditioning Systems Technical Manual Revision 000 Page 11 of 16 Appendix B Drawings B1 4S M8 flow diagram B2 Impinger schematic B3 Relay board contact state B4 Relay board...

Page 12: ...ecommended 3 The service diaphragm is secured by the single screw in its center Remove this screw with a 5 32 Allen wrench The diaphragm and its clamping plate should be easily lifted off Some slight...

Page 13: ...2008 ALL RIGHTS RESERVED PERMA PURE LLC SE MAN 003 Sample Conditioning Systems Technical Manual Revision 000 Page 13 of 16 Appendix D Dual Stream Operation The dual stream option was not purchased fo...

Page 14: ...9 0 469 0 431 0 396 0 363 0 333 40 35 30 29 28 104 95 86 84 82 7 25 5 55 4 19 3 95 3 73 8 9 10 11 12 18 16 14 12 10 0 305 0 281 0 256 0 234 0 214 27 26 25 24 23 81 79 77 75 73 3 62 3 32 3 13 2 94 2 77...

Page 15: ...48 3 79 62 80 00614 8 08 0328 5 01 60 76 00808 10 6 043 6 59 58 72 0106 13 9 0565 8 63 56 69 0138 18 2 0735 11 3 54 65 0178 23 4 0948 14 5 52 62 0230 30 3 123 18 8 50 58 0295 38 8 157 24 1 48 54 0378...

Page 16: ...in fact be suitable or desirable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing an...

Reviews: