Installation
Mounting
1.
Rack
2.
Wall
Wiring
Plumbing
Utilities
Start-up Instructions
LED Summary (see Cooler Manual)
Start-up
The gas sample pump is controlled and will operate only under certain conditions. The sample
pump will only be allowed to run if the temperature is 10°C or less AND the water slip sensor
indicators are green.
The peristaltic drain pumps will run continuously anytime Sample Conditioning System is
energized.
Now that there is a complete understanding of all of the controls and indicators, please verify that
all of the connections to the Sample Conditioning System are correct. Once this re-check has been
completed, please energize Sample Conditioning System. The unit should be ready to operate and
begin the sample conditioning process after operating for about ten minutes.
1.
Plug power cord into a properly grounded main circuit. Green “Slip” LED will come on.
2.
Wait for the green “Ready” LED light to come on, approx. 3 minutes. This will indicate the
ready temperature (+10°C) has been achieved and the sample gas flow can begin.
3.
After approx. 3 more minutes, the factory set-point of +5°C (41°F) or less will be reached.
The SLIP Green LED is always on unless;
Moisture is detected by the water slip sensor
The cooler was ordered without a relay board
There is a malfunction (eg. Shorted water slip sensor leads or a bad relay board)