Halma PERMA PURE Baldwin 4S-9PA Technical Manual Download Page 12

 
 
 

PERMA PURE LLC

Sample Conditioning 

Systems User’s Manual 
Doc. #580: Revision: 000 

Page 12 of 16 

 

Appendix C:  Perma Pure Sample Pump Maintenance Instructions 
 

Disassembly of head section and service diaphragm 
 

1.

 

Remove head section by unscrewing the four large bolts. A flat-bladed screw driver may be needed to 
gently pry the head free of the service diaphragm. If you have FEP coating on the heads, use caution 

not to scratch the surface. 

2.

 

The valve body can then be removed by unscrewing the two smaller screws (also accessible on the top 
of the head section). This part may be freed by gently tapping on these two screws after they have 

been loosened about 3 or 4 turns. When the valve body is removed, check all internal surfaces fro an 

accumulation of dirt. The two valve discs can be wiped clean and replaced as long as they appear 
unaffected by usage. The valve gasket can be easily removed and should be inspected. As a matter of 

good practice, the valve discs and valve gasket should be replaced during any routine maintenance 

check of the head section. A once a year routing procedure is recommended. 

3.

 

The service diaphragm is secured by the single screw in its center. Remove this screw with a 5/32” 
Allen wrench. The diaphragm and its clamping plate should be easily lifted off. Some slight adherence 

to the metal may occur if the diaphragm has been in use for a long period. 

4.

 

When replacing the service diaphragm, a FEP washer should be inserted under the head of the 
diaphragm cap screw. This is added insurance against small gas leaks through screw heads and may 

be essential in vacuum applications where outside air contamination cannot be tolerated. After 

tightening the screw, the excess FEP should be trimmed away. 

5.

 

NOTE: When replacing the service diaphragm, be sure the four projection studs of the base casting 

are properly located in the four outer holes provided in the diaphragm before the part is clamped in 

place. Be sure the diaphragm plate is firmly replaced with its center screw. 

 
Replacement parts: 

2PAM-002 

Sample pump repair kit, dual head 

Item 

Description 

 



10 

11 

12 
13 

14 

15 

16 
17 

18 

19 
20 

21 

22 
23 

24 

25 

VALVEBODY 

BEARING 
CAP 

CWEIGHT 

CAPACITOR 

ECCENTRIC 
FAN, BLUE 

FAN, WHITE 

HEAD STD, SS 
PLATE DIAPH. 

CONNECTING ROD 

TOL. RING 
SCREW-HEAD/BASE 

SCREW 

PLATE SCREW 

GROUND SCREW 
SCREW 

SCREW 

SCREEN 
SCREW 

LOCK WASHER 

DISC-VALVE 
DIAPHRAGM 

GASKET 

WASHER 

Summary of Contents for PERMA PURE Baldwin 4S-9PA

Page 1: ...A PURE LLC 732 244 0010 800 337 3762 fax 732 244 8140 www permapure com info permapure com PERMA PURE LLC 8 Executive Drive Toms River NJ 08755 Baldwin Series Sample Conditioning Systems Technical Man...

Page 2: ...nnections Error Bookmark not defined LED Summary 5 Start up 5 Set Points Error Bookmark not defined Shutdown Error Bookmark not defined Replacement Parts 8 Appendix A Installation of Peristaltic Pump...

Page 3: ...0410 Sample gas flow range Water content Maximum cooling rate Maximum inlet sample temp Maximum inlet pressure Maximum pressure drop Ambient temperature range Outlet sample gas dew point Voltage Power...

Page 4: ...PERMA PURE LLC Sample Conditioning Systems User s Manual Doc 580 Revision 000 Page 4 of 16 o Copy M Class Safety info...

Page 5: ...cators please verify that all of the connections to the Sample Conditioning System are correct Once this re check has been completed please energize Sample Conditioning System The unit should be ready...

Page 6: ...indicates ambient temperature Thermocouple placement in heat exchanger block Ensure proper placement Replace thermocouple or control board Impinger does not reach set temperature but is below ready te...

Page 7: ...use excess condensate and could lead to water slip Sample pump does not start Ready LED is on Slip LED is Off alarm relay water slip option installed Electrical connections to sample pump Verify that...

Page 8: ...3 Thermocouple Temperature Control Type K 36 3KFA 001 Complete filter assembly sample in line 2 micron 3FHG 001 Filter bowl glass 3FEC 002 Filter element ceramic 2 micron 3KPB 003 Peristaltic pump dua...

Page 9: ...not snap tightly together reload tubing If necessary turn key in slot on rotor shaft to adjust tubing as in step e e With key in slot on rotor shaft turn key to align tang on rotor shaft with slot in...

Page 10: ...PERMA PURE LLC Sample Conditioning Systems User s Manual Doc 580 Revision 000 Page 10 of 16...

Page 11: ...PURE LLC SE MAN 003 Sample Conditioning Systems Technical Manual Revision 000 Page 11 of 16 Appendix B Drawings B1 4S M8 flow diagram B2 Impinger schematic B3 Relay board contact state B4 Relay board...

Page 12: ...ecommended 3 The service diaphragm is secured by the single screw in its center Remove this screw with a 5 32 Allen wrench The diaphragm and its clamping plate should be easily lifted off Some slight...

Page 13: ...2008 ALL RIGHTS RESERVED PERMA PURE LLC SE MAN 003 Sample Conditioning Systems Technical Manual Revision 000 Page 13 of 16 Appendix D Dual Stream Operation The dual stream option was not purchased fo...

Page 14: ...9 0 469 0 431 0 396 0 363 0 333 40 35 30 29 28 104 95 86 84 82 7 25 5 55 4 19 3 95 3 73 8 9 10 11 12 18 16 14 12 10 0 305 0 281 0 256 0 234 0 214 27 26 25 24 23 81 79 77 75 73 3 62 3 32 3 13 2 94 2 77...

Page 15: ...48 3 79 62 80 00614 8 08 0328 5 01 60 76 00808 10 6 043 6 59 58 72 0106 13 9 0565 8 63 56 69 0138 18 2 0735 11 3 54 65 0178 23 4 0948 14 5 52 62 0230 30 3 123 18 8 50 58 0295 38 8 157 24 1 48 54 0378...

Page 16: ...in fact be suitable or desirable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing an...

Reviews: