2: Installing the Maya2000 Series
10
020-00000-000-02-201603b
Notes
Make sure the ALT checkbox for the GPIO (Index 4) is not selected, which would
prevent the shutter from triggering.
The activation time of the internal shutter is 11 ms.
Connect Spectroscopic Accessories
To find operating instructions for Maya2000Pro Series-compatible products (such as light sources,
sampling chambers, and probes), consult the Ocean Optics website at
http://www.oceanoptics.com/technical/operatinginstructions.asp
Change the Slit
There’s no need to calibrate your spectrometer when changing the slit, just install and start measuring.
There is one exception to this. You cannot change from a standard slit to a slit with a filter because it
changes the optical focus and wavelength calibration of the spectrometer. In this case you would need to
send the spectrometer back to Ocean Optics. A filter must be ordered for each slit (if your application
requires a filter) and the spectrometer needs to be calibrated and focused with the filter installed. This
only applies to filters installed inside the slit assembly.
►
Procedure
1.
Find the SMA connector. If a fiber is attached, remove it.