background image

8

6. Operation

 (cont'd)

External output

The terminal block for external output is located on the left side of the body. Pass/Fail results can be 
output from this terminal.
The following shows the output circuit. Output can be enabled by either "PASS" or "FAIL" according to 
the DIP switch settings.
As external output is photocoupler output, add a drive circuit when a large current is required for 
connection of a relay or other device.

1

2
3

FG-465 side

+

 terminal

-

 terminal

Output time: ON for 1 second after evaluation

Photocoupler output 
Max. absolute rating DC 80 V 
Max. output current DC 5 mA

DIP switch No.6 
OFF: External output on PASS 
ON: External output on FAIL

No.1, 2 

For external 

output

No.3 
For GND

Terminal block for external output (conditions of use)

Model

XW4C-03E1-H1 (made by OMRON)

Stripped length of lead

10 mm

Applicable lead

Single wire

0.2 mm

— 4.0 mm

2

Strand wire

0.2 mm

— 2.5 mm

2

Rod 
terminal

w/o sleeve

0.25 mm

— 2.5 mm

2

w/ sleeve

0.25 mm

— 1.5 mm

2

AWG

24 — 12

Connection to terminal block

Example: When directly connecting lead

(1)  Use a lead of size that falls under the applicable lead 

category.

(2)  Strip back 10 mm of the sheath from the end of the lead, 

and twist the end.

10 mm

 CAUTION

A correct connection is no longer possible if the end of the 
lead is pre-soldered.

(3)  Push in a screwdriver or similar tool into the release hole 

on the terminal block. (a)

(a)

(b)

Terminal insertion 
hole

Release hole

(4)  Insert the lead with sheath stripped back from its end 

into the terminal insertion hole. (b) 

When removing the lead, push a screwdriver or similar 

tool into the release hole, and draw out the lead.

Summary of Contents for FG-465

Page 1: ...ontents HAKKO FG 465 1 Installation base 1 AC adapter 1 Test plate 1 Knobs 2 USB cable 1 Pan head screws M5 x 10 2 Wood screws DN 4 5 x 16 2 Instruction Manual this document 1 HAKKO FG 465 Wood screws...

Page 2: ...85 D mm 4 7 1 2 7 3 in Weight 0 55 kg 1 2 lb including installation base Test plate Cord length 1 6 m 5 2 ft Dimensions w o cord 300 W 30 H 300 D mm 11 8 1 2 11 8 in Weight w o cord 1 9 kg 4 2 lb AC a...

Page 3: ...specialist Current flowing through human body 2 to 3 A approx reference value measurement on the antistatic shoes 20 M Current flowing through human body 50 to 100 A measurement on bare feet Wearing t...

Page 4: ...e UP Connection cord UP mark at toes Left foot Right foot See External output P 8 UP Power switch Connection cord Test plate left foot Test plate right foot Pad Pass range indicator Connection port fo...

Page 5: ...Insert the connection cord from the test plate and AC adapter plug into the jacks on the left side of the body 4 Insert the AC adapter into the power outlet CAUTION When mounting this product on a wal...

Page 6: ...ge Note Measurement of the right foot starts from the range finalized by measurement of the left foot If there is no problem with the footwear the range movement becomes minimum This results in a shor...

Page 7: ...M DIP switch CAUTION Before changing the DIP switch settings be sure to turn the power OFF DIP switch setting table Setting DIP switch No 8 7 6 5 4 3 2 1 Electrical Conductivity R 0 1 M ON ON Lower l...

Page 8: ...Terminal block for external output conditions of use Model XW4C 03E1 H1 made by OMRON Stripped length of lead 10 mm Applicable lead Single wire 0 2 mm2 4 0 mm2 Strand wire 0 2 mm2 2 5 mm2 Rod terminal...

Page 9: ...he FG 465 ON When these are connected correctly the buzzer sounds twice When the power is turned ON on just the FG 465 the buzzer sounds once During measurement the LED on the FG 470 changes state Thi...

Page 10: ...tch setting table The buzzer just sounds 3 times and the pass fail result is not output Did you move your hand away from the pad quickly Do no move your hand away from the pad until PASS or FAIL is fi...

Page 11: ...adapter CH plug B5119 AC adapter BS plug B5220 AC adapter European plug CE B3528 AC adapter 2 Core Flat pin plug 4 B1991 Knobs qty 2 5 B5262 USB cable 1 m Mounting example a c b Options Dwg No Part No...

Page 12: ...GAPORE HAKKO PRODUCTS PTE LTD TEL 6748 2277 FAX 6744 0033 https www hakko com sg E mail sales hakko com sg HONG KONG HAKKO DEVELOPMENT CO LTD TEL 2811 5588 FAX 2590 0217 https www hakko com cn E mail...

Reviews: