background image

9

 

Data communication function (DATA I/O)

Connecting an exclusive cable adaptor module (option) to the DATA I/O connector on the 
upper part of the HAKKO FG-101B enables communication to applicable stations such as the 
HAKKO FN-1010, etc. (See the figure below) Data communication by infrared will be disabled 
during communication by DATA I/O. Pressing the 

 button for 1 second or longer sends 

data. Only fixed values can be sent.
  In order to send fixed values, the AUTO HOLD function or MAX HOLD function can be used 
for temperature measurements and the AUTO HOLD function can be used for voltage or 
resistance measurements. Measurement and data acquisition can be performed smoothly 
using the functions of connected stations.

Ground terminal (GND)

When using a type of soldering iron that is grounded with an alligator clip, connect the clip to 
the GND terminal.

Maintenance and Calibration

• To replace the conduction plate, remove the set screws.
•  The life of the temperature sensor will vary depending on the temperature at which 

measurements are made and the type of solder and flux being used. In general, temperature 
sensors can be used for 50 measurements. Replace the sensor as soon as the measuring 
point wears out.

•  HAKKO can calibrate the instrument for a nominal fee. Please contact your dealer for further 

information.

 

CAUTION

Pressing the POWER button while holding the 

 button pressed changes the 

temperature unit between ˚C and ˚F.

 DIN cable

 RS232C cable

C5055

Adapter module

HAKKO

FN-1010

HAKKO

FG-101B

5.  MAINTENANCE

Summary of Contents for FG-101B

Page 1: ...a soldering iron tester capable of measuring soldering tip temperature leak voltage and ground resistance Please read the manual carefully before operating the HAKKO FG 101B Please keep this manual r...

Page 2: ...TIONS AND NOTES 3 4 OPERATION 4 4 1 Initial setup 4 4 2 Measurement of tip temperature 5 4 3 Measurement of leak voltage 6 4 4 Measurement of resistance difference between tip and ground wire 7 4 5 Ex...

Page 3: ...isplay contents 1 This alarm indicates sensor burnout If this alarm occurs replace the sensor 2 When the and buttons are held pressed simultaneously for a long time 1 sec or longer the count is reset...

Page 4: ...nctions Temperature measurement count function 000 to 999 times Power consumption 100V 3 2W 100 110V 3 6W 120V 3 2W 220 240V 3 6W Outline dimensions 211 W 53 H 126 D mm Weight 0 95 kg Usable temperatu...

Page 5: ...rning use of the appliance in safe way and understand the hazards involved Children shall not play with the appliance Cleaning and user maintenance shall not be made by children without supervision Be...

Page 6: ...will move toward the terminal 2 Attach the sensor with the slide pin moved toward the terminal 3 Attach the sensor having a red mark to the red colored terminal and the sensor having a blue mark to th...

Page 7: ...When flux is stuck on the terminal wipe it off with alcohol Do not use thinner or benzin for wiping Read the values when the display temperature is stable CAUTION Do not measure the temperature by ap...

Page 8: ...ed into the power receptacle the HAKKO FG 101B main body 2 Wait until the tip reaches the set temperature If the soldering iron temperature is variable set temperature to maximum 3 Press the SELECT bu...

Page 9: ...the button will be displayed Pressed for a long time 1 second or longer The MAX HOLD function will be released and the product will return to normal display CAUTION Turning off the power always cance...

Page 10: ...Each time the button is pressed the AUTO HOLD function toggles between ON and OFF The AUTO HOLD function does not assume the measurement by temperature probes such as the hot air Data send function in...

Page 11: ...performed smoothly using the functions of connected stations Ground terminal GND When using a type of soldering iron that is grounded with an alligator clip connect the clip to the GND terminal Maint...

Page 12: ...nese plug B3508 Power cord 3 core American plug B3550 Power cord 3 core SI plug B3616 Power cord 3 core BR plug B1752 Conduction Plate B1754 Ground Clip B1950 Conduction Wire B1258 Fuse 250V 3 15A B24...

Reviews: