Hakko Electric Machine Works Co., Ltd. HAP2031 Handling Manual Download Page 19

18

《HOW TO SET FREQUENCY 》 

Indication is shown for HAP2000Fseries and HAP2000T series、for
HAP2000 series but  it is not effective due to without inverter.

CAUTION

1.  Push "mode changing key" one time in operation mode.

Present value display goes into "

" and into fre-

quency setting mode.

2. Pushing "data changing key"

and change target value

of setting value display. Range for setting is 30- 60[Hz].

After 5 seconds, going on and off will cease and
fix. it is also possible to fix value with pushing
"mode changing key". ( in this case, it goes into
timer setting mode)

℃ 

時.分 

Hz

℃ 

熱風温度(PV) 

温度設定(SV)/周波数/時間 

code for frequency
setting mode.

frequency setting
value.

℃ 

時.分 

Hz

℃ 

熱風温度(PV) 

温度設定(SV)/周波数/時間 

setting range of 30-60
going on and off dur-
ing changing.

《HOW TO SET TIMER》

This is setting methods of timer which counts with
"timer operation mode" and "timer stop mode".

1.  Push twice mode changing key at operation mode.

Present value display changes into "

" and into

timer setting mode.

2. Pushing "data changing key"

and change target value

of setting value display. Range for setting is 00hr 00min 〜
99 hrs 59min. 

After 5 seconds, going on and off will cease and
fix. it is also possible to fix value with pushing
"mode changing key". (hence, into remaining time
of timer monitor mode)

℃ 

時.分 

Hz

℃ 

熱風温度(PV) 

温度設定(SV)/周波数/時間 

code for timer setting
mode.

timer setting value.

℃ 

時.分 

Hz

℃ 

熱風温度(PV) 

温度設定(SV)/周波数/時間 

lowest two digits : minutes
upper two digits : hours

range of setting 0000 〜
9959. going on and off
during changing.

If setting is made in 00hrs00min, timer becomes not effective.
Operation key for timer and stop key for timer also become not effective.

CAUTION

During timer operation or stopping, if time for setting is changed, value
set later will be effective.

CAUTION

《HOW TO CHECK REMAINING TIME OF TIMER 》

1.  Push "mode changing key" three times at operation mode.

Present  value  display  changes  into  "

"  and

thereafter  into  checking  mode  of  remaining  time
of  timer.  Remaining  time  of  timer  is  indicated  in
setting display.

℃ 

時.分 

Hz

℃ 

熱風温度(PV) 

温度設定(SV)/周波数/時間 

lowest two digits : minutes
upper two digits : hours

code for remaining
time of timer monitor
mode.

remaining time of
taimer.

Push "mode changing key" two times
in monitor of remaining time of timer

and it will go into operation mode, but 5 min after without
any operation it goes operation mode automatically.

CAUTION

《HOW TO SET FREQUENCY 》《HOW TO SET TIMER 》

《HOW TO CHECK REMAINING TIME OF TIMER 》

During timer operation  or stopping,  it is possible to check remaining time of timer.

Summary of Contents for HAP2031

Page 1: ...hasing Please follow instructions herein thoroughly for correct use Hold this manual at convenient place for reminding Please check the followings when goods arrived The name plate of the goods if all...

Page 2: ...RATION WIND FAN OPERATION FAN OPERATION 15 STOP 15 TIMER OPERATION 16 TIMER STOP 16 CONTROLLER OPERATION METHOD 17 HOW TO SET CONTROL TEMPERATURE 17 HOW TO SET A FREQUENCY 18 HOW TO SET TIMER 18 HOW T...

Page 3: ...are not handled properly might become cause of death incident or fire It is difficult for the company to imagine all about potential risks which might occur in the future but every probability of ris...

Page 4: ...and outlet during operation Do not close inlet and outlet during operation It will cause of firing and out of order Do not operate by wet hand Do not operate by wet hand It will cause of electric shoc...

Page 5: ...age Do not use others than indicated power and voltage It will cause of fire electric shock out of order Do not put flammable things around inlet It will cause of fire or out of order if it should be...

Page 6: ...5 3m 4core 5 5 3m 4core 8 3m Weight 28 kg 34 kg 35 kg Model No HAP2151 HAP2201 HAP2301 Product Code 00700260 00700270 00700280 Power Source 3 phase 200V 50 60Hz Total Wattage 16 5 kW 21 5 kW 31 5 kW H...

Page 7: ...kg 35 kg Model No HAP2151F HAP2201F HAP2301F HAP2402F Product Code 00700261 00700271 00700281 00700293 Power Source 3 phase 200V 50 60Hz Total Wattage 16 5 kW 21 5 kW 31 5 kW 41 5 kW Heater Wattage 1...

Page 8: ...e at Max Air Flow 80 dB 89 dB Power source electric wire 4core 3 5 3m 4core 8 3m Weight 42 kg 72 kg HAP2000T SERIES Air temperature varies depend on usage conditions Use with conditions of under maxim...

Page 9: ...703 558 45 313 170 55 280 395 190 423 732 4 7 14 round long hole fixing hole 73 70 200 480 75 250 400 523 1016 818 280 650 1050 60 70 148 123 135 Hose joint pipe F siries only fixing hole 4 12 24 roun...

Page 10: ...Fixing Bracket Fan Motor Inlett for Cool Wind Outlet Bracket for Fixing Outlet Panel Lifting Bolt Name Plate Handle Not available for HAP2051T Inlet Power Breaker main body HAP2051T HAP2101T MAIN BOD...

Page 11: ...ring timer being stopping Delay Timer Lamp It lights on and off during delay timer being counting Outer Control Lamp It lights on when input of outer control outer wind fan outer hot wind fan and oute...

Page 12: ...076 HAP2081 HAP2101 HAP2031F HAP2051F HAP2076F HAP2081F HAP2101F M6 2 p ce HAP2051T M6 4 p ce HAP2151 HAP2201 HAP2151F HAP2201F HAP2101T M8 4 p ce HAP2301 HAP2402 HAP2301F HAP2402F M10 4 p ce 1 It is...

Page 13: ...amper with flange by 4 fixing bolts free band and flexible hose are optional parts Free Band Out let Flexible Hose Packing Damper with Flange Free band Fixing Bolts Flexible Hose CAUTION Select and us...

Page 14: ...reaker of installation Those are R red white black and earth green Fasten tightly with solders ring terminals type length product code ZAA1103 3 m 00950515 ZAA1105 5 m 00950525 ZAA1110 10 m 00950535 c...

Page 15: ...g delay timer counting delay timer lamp and wind fan lamp put out lights and wind fan stops TRIAL OPERATION ADJUSTMENT In case of reverse rotation of wind fan maximum wind vol ume and static pressure...

Page 16: ...ts off delay timer lamp and wind fan lamp wind fan stops It puts off power lamp of controller During delay timer lamp is going on and off wind fan operates for the time 2 minutes in delay timer counti...

Page 17: ...how to set CAUTION To check remaining time of timer During timer operation or stopping it is possible to check remaining time of timer Refer to page 18 on the detail of how to check TIMER OPERATION TI...

Page 18: ...o HAP controller handling manuals details division It is available from h p http www hakko co jp for HAP controller handling manuals details division HOW TO SET CONTROL TEMPERATURE 1 Put on a breaker...

Page 19: ...d fix it is also possible to fix value with pushing mode changing key hence into remaining time of timer monitor mode Hz PV SV code for timer setting mode timer setting value Hz PV SV lowest two digit...

Page 20: ...N closure contact in irregularity AC100 200V 1A resistant road 0ver heat out put Contact out put becomes ON in case that liquid expansion type thermo stat equipped in a main device detects over heat i...

Page 21: ...e irregularity by thermocouples connected with in put of thermocouples 2 at outer control terminal By this indication of code it puts off to heater and wind fan operation begins with starting delay ti...

Page 22: ...and etc Check outlet of furnace and etc High temperature of suction air Setting error of controller parameter Get suction air temperature lower mixing fresh air Check parameter setting with list of co...

Page 23: ...neutral gender detergent 1 In case of stocking in as being packed Stock indoor in less temperature difference and dry humidity Do not stock in double accumulation 2 In case of stock as being installed...

Page 24: ...and terminal of outer control wind fan hot fan COM respectively Hz PV SV outer control input effective not effective setting code invalid with OFF Hz PV SV top of parameter block 1 mode Pushing mode...

Page 25: ...ut terminal and outer temperature control terminal to out put of SSR control of temperature controller DC12V pulse out put light on hot wind outer control Pushing mode change key long time 1 second at...

Page 26: ...r mocouples control with 1 of thermocouples equipped in device Hz PV SV control with 2 of thermocouples to in put of thermocou ples 2 of outer con trol terminal 1 It is designated that a sensor for te...

Page 27: ...ode change key each one time in order set value display is changed in and is fixed Control is set with thermocouples wired with in put of thermocouples 2 of outer control termi nal Hz PV SV alter sett...

Page 28: ...ndicate irregularity in reaching at temperature for working Hz PV SV alter setting value non indication with oFF Hz PV SV indication of temperature control irregularity 1 setting corresponding with th...

Page 29: ...ce Push key one time in mode It changes for present value display to and for set value display to It goes to set mode of working temperature of temperature control irregularity 2 Set value to with pus...

Page 30: ...ication dSP4 1 I 120 indication dSP4 2 d 30 indication dSP4 4 HYS 2 not indication dSP4 8 bAL 0 not indication dSP4 16 Ar 100 not indication dSP4 32 TC 2 indication dSP4 64 P1n2 3 not indication dSP4...

Page 31: ...2 38 indication dSP3 0 indication dSP4 184 indication dSP5 255 indication dSP6 1 indication dSP7 255 indication dSP8 14 indication blind mode brM br1 on br2 on br3 on Please refer to this list when it...

Page 32: ...17 Tomeoka Miyaginoku Sendai City TEL 022 257 8501 FAX 022 257 8503 SAPPORO SALES OFFICE 060 0004 Sankyo Garden Heights West15 1F 15 1 35 Kita SijyouNishi Chuoku Sapporo City TEL 011 611 8580 FAX 011...

Reviews: