-------Internal training document for Haier service provider. External disclosure is prohibited --- Classification
13
13
temperature is ≥13
℃
(for 5min. continuously), the “hot condenser” indicator turns on and the machine
makes the relevant silence alarm. After the power supply is on, the indicator does not turn on and the
machine does not make a silence alarm in 3h.
2)
When the sensor has a failure, the indicator of the “sensor failure” turns on;
3)
When the main sensor has a failure, the failure code “E2” of the main sensor and the actual
temperature are displayed alternatively on the display panel.
4)
When the condenser sensor has a failure, the failure code “E1” of the main sensor and the actual
temperature are displayed alternatively on the display panel.
5)
When the ambient temperature sensor has a failure, the failure code “E0” of the main sensor and the
actual temperature are displayed alternatively on the display panel.
6) When the heat exchanger
sensor has a failure, the actual temperature and the failure code “E3” of the
heat exchanger sensor display alternatively on the display panel. When you select -
86 and “E3” appears,
after you press the button for clearing the silence alarm for 5s, “E3” disappears. After “E3” displays, after
the high temperature compressor is started, the low temperature compressor will start with a 1min. delay.
7) When the temperature detected by the heat exchanger sensor is
≥90
℃
, the alarm displays E4.
Inward main sensor E2
Ambient
temperature sensor
failure E0
The heat exchanger
probe foams at the
bottom
inside
the
freezer.
E3 at the terminal of
the heat exchanger
sensor
Condenser sensor E1