17
User Manual
Precision of power setting radio frequency (HF)
The power of the radio frequency generator depends on the tissue specific resistance and can vary within limits. The specified
50 W are in accordance with a specific resistance of 1 kΩ. The settings of the radiofrequency will be displayed without the
measuring unit [W] and will be scaled in accordance with the graph.
Meyer-Haake Innovations GmbH / Ahlmeier GmbH
18.07.2017
0
50
100
150
200
250
300
350
0
10
20
30
40
50
U/V
Leistungsanzeige
Max. Ausgangsspannung in Betriebsart "Cut"
Max. output voltage in operating mode „Cut“
Performance indication
0
10
20
30
40
50
60
0
500
1000
1500
2000
P/W
R/
W
HF-Leistung in Betriebsart "Cut" vs. Last
25
50
RF-power in operating mode „Cut“ vs. load
0
10
20
30
40
0
500
1000
1500
2000
P/W
R/
W
HF-Leistung vs. Last in Betriebsart "Cut/Coag" bzw. "Coag"
bei verschiedenen Leistungseinstellungen, Koagulationsgrad C4
15
30
RF-power vs. load in operating mode „Cut/Coag“ or „Coag“
in various power settings, coagulation level C4
0
50
100
150
200
250
300
350
0
10
20
30
U/V
Leistungsanzeige
Max. Ausgangsspannung vs. Anzeige in Betriebsart
"Cut/Coag" bzw. "Coag", Koagulationsgrad C4
Performance indication
Max.output voltage vs. visual display in operating mode
„Cut/Coag“ or „Coag“, coagulation level C4