User Manual

The hf Surg device may only be put into operation after the instructions have been given by the operator and in compliance with the 
prescriptions and safety provisions.

Protection and safety provisions in the dental office for the use of the hf Surg device

WARNING

The device must only be connected to a power supply with an earth wire in order to avoid the risk of an electric shock. 

Check the cables, hand piece and electrodes as well as the foot pedal on visible damage before starting up 

the hf Surg device. Instruments with brittle or faulty insulation must not be used as they pose a danger of 

injury. 

When operating the hf Surg device, unpredictable malfunctions might occur that could cause un-

wanted output power increase.

Technical Data

HF

fre power:

2,2 MHz, max. 50 W, load resistance 1000 Ohm

method of operation

monopolar

wave form

permanently / pulsed with ca. 80 %

Ambient temperature

+5 °C – +40 °C

Storage temperature

-20 °C – +70 °C

Humidity

Operating humidity: < 85%, non-condensing
Storage humidity: < 90%, non-condensing

Air pressure

during storage : 500 hPa - 1080 hPa

during operation : 700 hPa - 1080 hPa

Power supply

230 V AC, 50/60 Hz

Power consumption

max. 100 VA

Medical device classification

class IIb

Applicators

Hand piece

autoclavable

Safety

According to DIN EN ISO 6060-1

Electromagnetic compatibility

According to DIN EN ISO 6060-1-2

Safety hf surgery equipment 

According to DIN EN ISO 60601-2-2

HF handpiece cable, yellow

Length 1.50m

HF neutral electrode cable

Length 1.50m

Foot switch cable

Length 2.50m

Power cable

Length 2.50m

Cables

Summary of Contents for 452400

Page 1: ...HF Chirurgieger t Bedienungsanleitung HF surgery unit User Manual hf SURG...

Page 2: ...wesentlichen Leistungsmerkmale nach EN 60601 1 zugeordnet Anwendungsbereiche Das Einsatzgebiet sind professionelle Einrichtungen in der Zahnheilkunde nderungen Der Hersteller beh lt sich das Recht vo...

Page 3: ...einen Stromschlag erleidet wird ein hochfrequenter Strom verwendet THERMISCHE WIRKUNG MONO POLAR SCHNEIDEN CUT KOAGULIEREN COAG Bei dem Vorgang des monopolaren Schneidens wird der hochfrequente Strom...

Page 4: ...tromagnetische Vertr glichkeit Gem DIN EN ISO 60601 1 2 Sicherheit HF Chirurgie Gerate Gem DIN EN ISO 60601 2 2 Schutz und Sicherheitsbestimmungen in der zahn rztlichen Praxis bei der Benutzung des hf...

Page 5: ...ttmachern oder anderen aktiven Implantaten besteht eine m gliche GEF HRDUNG dadurch dass eine St rung der Schrittmacherfunktion auftreten kann oder der Schrittmacher besch digt werden k nnte Im Zweife...

Page 6: ...oder schnei den koagulieren wollen Au erbetriebnahme nach Anwendung 1 Handst ck in die Ablage stecken 2 Elektrode entfernen und entsprechend aufbereiten 3 Ger t mit dem Hauptschalter 1 R ckseite auss...

Page 7: ...lung Defibrillationsgesch tztes Anwendungsteil Herstelldatum 6450300004 1252 D001 452400 Ger tevorderseite 1 Regler 2 Ausgang Handst ck 3 Ausgang Neutralelektrode 4 Wahltaste 5 Schneidmodus 6 Schneid...

Page 8: ...g in Betriebsart Cut vs Last 25 50 0 10 20 30 40 0 500 1000 1500 2000 P W R W HF Leistung vs Last in Betriebsart Cut Coag bzw Coag bei verschiedenen Leistungseinstellungen Koagulationsgrad C4 15 30 HF...

Page 9: ...nis Sie werden bemerken dass die Intensit tseinstellung zu hoch gewesen ist wodurch Funkenschlag und bemerkenswerte Verf rbung entlang der Schneidspur verursacht wurden 7 ReduzierenSiedieIntensit tauf...

Page 10: ...it Reinigungst chern und geeigneten Reinigungsmitteln s ubern F r die maschinelle Rei nigung ist der schraubbare Kopf vom Handst ck zu entfernen und dieser in das entsprechende Ger t zu legen Handst c...

Page 11: ...bility by Hager Werken GmbH Co KG is given if installation and start of operation is done by own personal or by personal authorized by the manufacturer installation and safety measures comply with nat...

Page 12: ...ION MONOPOLAR CUTTING CUT COAGULATION COAG At monopolar cutting the radio frequency current is led from the device via a work electrode and a large neutral electrode back to the device The current den...

Page 13: ...tions might occur that could cause un wanted output power increase Technical Data HF frequency power 2 2 MHz max 50 W load resistance 1000 Ohm method of operation monopolar wave form permanently pulse...

Page 14: ...the patient e g on the instrument table 5 The output power should be set as low as possible for the corresponding purpose 6 An obviously low output value or functional failure of the hf Surg device i...

Page 15: ...ant to cut or cut coagulate Lockout after use 1 Place the hand piece into the instrument holder 2 Remove the electrode and reprocess it accordingly 3 Switch off the device to position 0 using main swi...

Page 16: ...OAG 2 1 4 3 5 6 3 1 2 4 Unit Back 1 ON OFF switch 2 2230 V Power supply with fuses 3 Footswitch 4 Label Label Attheback side of the unit Explanations Manufacturer On Off Mains fuse Foot cotrol CE mark...

Page 17: ...rt Cut Max output voltage in operating mode Cut Performance indication 0 10 20 30 40 50 60 0 500 1000 1500 2000 P W R W HF Leistung in Betriebsart Cut vs Last 25 50 RF power in operating mode Cut vs l...

Page 18: ...also not suited Wait until the beef has obtainedroom temperature Make sure that the meat is placed on the plugged in neutralelectrode If this is not the case the waves cannot derivate and consequentl...

Page 19: ...Sight check on damages wear deformation Storage No special requirements The a m instructions have been validated as SUITABLE for preparation of a medical device and its re use by the medical device ma...

Page 20: ...12803 Rev1 2021 01 WEEE Reg Nr DE 21760541 HAGER WERKEN GmbH Co KG Ackerstra e 1 47269 Duisburg Germany T 49 203 99269 0 F 49 203 299283 info hagerwerken de www hagerwerken de Medizinprodukt Medical d...

Reviews: