![Hadeco Smartdop 30EX Operating Manual Download Page 54](http://html.mh-extra.com/html/hadeco/smartdop-30ex/smartdop-30ex_operating-manual_577371054.webp)
47
(11) Smartdop will automatically stop the measurement process, and then freeze the
waveform. Press PRINT button to print the waveform, if necessary.
(12) If you wish to save the data on the memory, see
“§ 4-2-7-a. MEMORY - STORE”.
(13) Press the INF/DUMP button to get out of the
freeze mode.
7-4-3. Menu for PV
Menu
Sub Menu
Selections
Reference in § 4-2-7
MEMORY
STORE*²
1 to 30, FREEZE
a. Memory - Store
READ
1 to 30, FREEZE
b. Memory - Read
CLEAR
1 to 30, ALL
c. Memory - Clear
IINFLATOR*¹
ON
, OFF
6-3-5. Inflator
OTHERS
LANGUAGE
ENGLISH
, DEUTSCH,
ITALIANO, ESPANOL,
FRANCAIS
k. Others - Language
PRB
– KEY
(Probe button)
FREEZE, PRINT, PRT & FRZ
l. Others - PRB-KEY
AUTO-OFF
ON
, OFF
p. Others - AUTO-OFF
DISCHARGE*³
q. Others - Discharge
Note
*1
: INFLATOR is available only on Measurement mode.
*2
: MEMORY - STORE is available only on Freeze mode.
*3
: DISCHARGE is available when AC adaptor is connected to the main unit.
7-4-4. PV Mode Setting
MODE (AC / DC) (Only Measurement mode);
(1) Press MODE button to select the mode. Mode LED corresponding to the mode will
illuminate.
Mode 1:
AC coupling mode for arterial pulse waveform studies
Mode 2:
DC coupling mode for measurement of maximum venous outflow
INFLATOR;
(1) Go to INFLATOR on MENU and press Enter or Right to change the mode as follows:
ON:
Inflator ON when used with PV-21 without sphyg.
OFF:
Inflator OFF when used with PV-21 with sphyg.
(2) When INFLATOR is ON, connect 3-way stop cock to inflation inlet of Smartdop instead
of sphyg with a tubing and Smartdop will inflate the cuff to 60 mmHg automatically.