45
7-4. Pneumoplethysmograph
With the PV-21, Smartdop senses volume changes in a limb or digit by measuring the
pressure changes in a recording cuff.
“How to use pneumoplethysmograph” is described in this section.
Built-in cuff inflator works when INFLATOR in menu mode is ON to inflate cuff.
See “§ 7-4-4. PV Mode Setting - INFLATOR” for details.
7-4-1. PV (Pneumoplethysmography) Probe Assemblies
Model PV-21
DC Coupling:
Measurement of maximum venous outflow
7-4-2. PV - Measurement of Maximum Venous Outflow
Purpose
Because deep venous occlusion cannot be diagnosed reliably on the basis of presenting
signs and symptoms, pain and swelling being frequently due to other causes, objective
screening tests are of value in confirming or ruling out suspected venous obstructions in
the lower extremities. The test consists of first inducing temporary venous pooling by
means of a constricting thigh cuff followed by measurement of the rapidity of emptying
when the constricting cuff is suddenly vented. Measurement of maximum venous outflow
is frequently employed as an adjunct to Doppler venous compression studies.
Preparation
(1) Connect the PV probe assembly (PV-21) the probe connector #1 or #2 of the main unit,
and turn it on.
(2) Make sure Mode 2 LED (DC mode) illuminates. If not, press MODE button to select
Mode 2 for DC mode.
To the main unit
To the cuff
Probe button