3-8. PV
– Arterial studies
(1) Wrap VC-12 cuff around patient's leg, thigh at
groin, above knee, below knee, or ankle.
(2) Connect the cuff to the tubing and other end of
tubing to the inflation inlet as shown in the right.
(3) Turn the unit on and press Probe Selection button
a few times until both #1 and #2 LEDs on the
button illuminate to go to PV-AC mode.
LCD will display as shown right below.
(4) Press INF/DUMP button to inflate the cuff up to 60
mmHg automatically.
Note: Press INF/DUMP button to stop the inflation and
dump cuff while inflation is in progress if desired.
The gain is automatically adjusted and the PV
waveform is shown on the LCD.
(5) When the waveform gets stable and rhythmic,
press INF/DUMP (Freeze) to freeze the waveform
and to dump the cuff.
Press PRINT button to print the waveform of last 5
sec, if necessary.
(6)
If you wish to save the data on the memory, see Ҥ
4-2-7-a. MEMORY -
STORE”.
(7) Remove the cuff and wrap it to the next site. Press INF/DUMP or probe button to
unfreeze waveform and repeat steps #5 to #7 for next site.
Inflation inlet